Success rate and proposed mechanism of balloon rupture in hemostasis with MYNX CONTROL.

Kaido Kobayashi, Kei Ichihashi, Kazuhiro Dan

一句话结论 · In one sentence

Balloon rupture was an infrequent but notable complication and occurred only in cases involving antegrade access and severe vascular calcification. Careful consideration of puncture direction and vascular calcification may further improve procedural safety when using MYNX CONTROL.

原始摘要(英文原文)· Original abstract
OBJECTIVES: Real-world data regarding the efficacy and safety of the MYNX CONTROL vascular closure device (Cordis) remain limited. The authors evaluated the efficacy and safety of MYNX CONTROL in routine clinical practice. METHODS: This single-center retrospective observational study included 127 consecutive patients who underwent femoral artery closure with MYNX CONTROL between August and December 2025. RESULTS: The primary endpoint was immediate hemostasis success, which was achieved in 120 patients (94.5%). Secondary endpoints included device-related complications, with a total of 7 events: balloon rupture (n = 2), minor bleeding (n = 2), hematoma (n = 2), and pseudoaneurysm (n = 1). Both cases of balloon rupture occurred during antegrade puncture in severely calcified vessels. CONCLUSIONS: Balloon rupture was an infrequent but notable complication and occurred only in cases involving antegrade access and severe vascular calcification. Careful consideration of puncture direction and vascular calcification may further improve procedural safety when using MYNX CONTROL.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

Success rate and proposed mechanism of balloon rupture in hemostasis with MYNX CONTROL. — 科研速览 Science Skim