Evaluating the Compressive Strength of Calcium Triglyceride-Based Polyurethane Bone Cement Composite: A Biodegradable Spacer Alternative.
Mohammed Ali Khashab
Department of Neurosurgery, Section of spine surgery, Ministry of the National Guard-Health Affairs; King Abdullah International Medical Research Center; King Saud bin Abdulaziz University for Health Sciences, Jeddah, Saudi Arabia.
Key points
From the paper's conclusion- Discrepancies exist between manufacturer-reported data and independent measurements. Although CTP bone cement demonstrated measurable compressive resistance, its observed performance suggests that it may be better suited to low-load-bearing applications. Independent validation is essential before in vivo or clinical use.
Abstract
OBJECTIVE: To independently test the compressive strength of calcium triglyceride-based polyurethane (CTP) bone cement under different curing conditions, assess its mechanical stability, compare the findings with manufacturer-reported data, and determine its potential for clinical applications. MATERIAL AND METHODS: This study tested the compressive strength of CTP bone cement using a mechanical compression machine in accordance with an American Society for Testing and Materials (ASTM) protocol. Samples were prepared according to the manufacturer's guidelines and tested after curing under various conditions: 135 minutes at ambient temperature, 24 hours at 37°C, and 2 months in water at 37°C. Testing was performed using tubular samples measuring 8 mm in height and 12 mm in diameter until the machine reached its 3500-N safety cutoff. A displacement of 4 mm was considered to represent 50% strain. RESULTS: CTP bone cement exhibited stable compressive resistance across all curing phases; the best performance was observed in samples cured at 37°C for 24 hours. The material maintained its compressive resistance after two months in water, demonstrating resistance to rapid degradation under near-physiological conditions. The highest load measured was 3500 N, which was significantly lower than the more than 100-MPa compressive stress reported in the manufacturer's data at comparable displacements. CONCLUSION: Discrepancies exist between manufacturer-reported data and independent measurements. Although CTP bone cement demonstrated measurable compressive resistance, its observed performance suggests that it may be better suited to low-load-bearing applications. Independent validation is essential before in vivo or clinical use.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Characterization of Water-Based Porous Calcium Phosphate Bone Cements for Peri-Implant Regeneration: An In Situ Study.Biomimetics (Basel, Switzerland) · 2026
- Micro-CT analysis of calcium silicate-based cements in simulated acidic and neutral environments: a comparative study.BMC oral health · 2026
- Redefining injectable PMMA bone cement: Beta-tricalcium phosphate integration for bioactive and osteoconductive performance for minimally invasive applications.Bio-medical materials and engineering · 2026
- Tailoring Segmented Polyurethane Foams for Bone Tissue Engineering: Effects of Hard Segment Content and Dual-Function Microhydroxyapatite.Materials (Basel, Switzerland) · 2026
- The degradation behavior of 3D-printed polyurethane composite scaffolds regulates bone regenerationBiofabrication · 2026
- Adhesive Bonding to Hydraulic Gel-Forming Calcium Silicate-Based Cements: Effects of Cement Type, Adhesive Strategy, and Restoration Timing.Gels (Basel, Switzerland) · 2026