Beyond Conventional RP-HPLC: Quality-by-Design, Green Analytical Chemistry, and Regulatory Perspectives for Multicomponent Analysis of Antihypertensive Drug Combinations.
Shaikh Tokir Arifhusen, Pratyush Kumar Brahma
Gujarat University
Key points
From the paper's conclusion- ICH Q14 and Q2(R2) provide a science- and risk-based framework for analytical procedure development and validation, whereas ICH M10 specifically addresses bioanalytical methods and study-sample analysis.
Abstract
Multicomponent analysis of antihypertensive drugs sits at the intersection of pharmaceutical quality control, bioanalysis, pharmacokinetics, and clinical therapeutic drug monitoring (TDM). Conventional reversed-phase high-performance liquid chromatography (RP-HPLC) remains useful because of its broad availability and flexible detection options; however, a method optimized for a synthetic mixture or tablet extract cannot automatically be transferred to plasma or serum, where endogenous components, concomitant medicines, protein binding, and low analyte concentrations can compromise selectivity and quantitative reliability. At the same time, method development is increasingly expected to demonstrate scientific understanding, robustness, lifecycle control, and reduced environmental burden. This review critically integrates analytical quality by design (AQbD), green analytical chemistry (GAC), chromatographic selectivity, sample preparation, detection, and regulatory expectations into a purpose-led framework for multicomponent antihypertensive analysis. Published bioanalytical examples demonstrate that LC-MS/MS can support targeted measurement of nebivolol in human plasma and simultaneous quantification of a broad panel of antihypertensive drugs in serum. These studies establish feasibility for their stated purposes, but they do not establish that one universal method is suitable for every drug, matrix, concentration range, or clinical question. ICH Q14 and Q2(R2) provide a science- and risk-based framework for analytical procedure development and validation, whereas ICH M10 specifically addresses bioanalytical methods and study-sample analysis. GAC metrics such as AGREE, GAPI, and Analytical Eco-Scale can make environmental trade-offs more transparent, although their outputs are not interchangeable and do not replace assessment of analytical fitness. The review proposes an integrated decision pathway linking intended use, analytical target profile, risk-based experimentation, matrixappropriate validation, sustainability assessment, and lifecycle monitoring. The principal research need is transparent comparative evidence that evaluates analytical performance and environmental burden together without sacrificing the reliability of pharmacokinetic or TDM conclusions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Integration of analytical quality by design green and white analytical chemistry in RP-HPLC method developmentDiscover Chemistry. · 2026
- Strategic integration of AQbD principles in RP-HPLC: a business perspective on drug quantification.Journal of chromatography. B, Analytical technologies in the biomedical and life sciences · 2026
- QbD-driven RP-HPLC method for the simultaneous analysis of dihydropyridines calcium channel blockers in pharmaceuticalsBMC Chemistry · 2025
- A Validated RP‐HPLC Method for the Concurrent Quantification of Two Amlodipine‐Based Ternary Mixtures With a Trio Color‐Coded Sustainability AssessmentSEPARATION SCIENCE PLUS · 2026
- A Validated Reversed Phase‐High Performance Liquid Chromatographic Method for Simultaneous Estimation of Bisoprolol Fumarate, Cilnidipine, and Telmisartan With RGB12‐Based Sustainability AssessmentSEPARATION SCIENCE PLUS · 2026
- Central Composite Design Approach Stability-Indicating RP-HPLC Analysis for Simultaneous Estimation of Monoclonal Antibodies in Bulk and Combined Dosage Form.Assay and drug development technologies · 2026