Monte Carlo study of noise-induced phase angle errors in low-frequency tetrapolar bioimpedance measurements.
Jie Hou, Sisay Mebre Abie, Ørjan G Martinsen
Department of Physics, University of Oslo, Oslo, Norway.
Key points
From the paper's conclusion- We used Monte Carlo simulations of lock-in detection to quantify how large phase errors that measurement noise alone can cause.
Abstract
Phase angles measured at low frequencies with the tetrapolar technique are sometimes scattered or even positive. We used Monte Carlo simulations of lock-in detection to quantify how large phase errors that measurement noise alone can cause. For a signal-to-noise ratio after demodulation (SNRout) of about 3 or higher, the standard deviation of the phase error was close to 57° divided by SNRout, and 95% of the errors were within about twice this value. At SNRout = 10 the error standard deviation was 6°; at SNRout = 1 it was 64° and 16% of the measurements were wrong by more than 90°. As SNRout decreased towards zero, the phase gradually approached a completely random value. In an illustrative low-frequency tetrapolar measurement, the absolute phase errors were small (standard deviation 0.007° at 1 Hz and 0.04° at 0.1 Hz), but the true tissue phase was even smaller. The error was therefore 1.2 times the true phase at 1 Hz and 37 times at 0.1 Hz, and the measured phase had the wrong sign in 20% and 49% of the measurements, respectively. The sign of a small phase angle is only reliable (wrong in less than 1% of measurements) when the true phase is larger than 2.3 times the error standard deviation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Maximum signal-to-noise ratio enhancement by averaging under a limited measurement timearXiv · 2026
- Examining within-session test-retest reliability of four bioimpedance devices among adults across a wide age range.Journal of electrical bioimpedance · 2026
- A 140.5$\mu$W, 99.8% Accuracy, 96.1dB SNR BioZ IC for Real-Time Cell Analysis with On-Demand Push-Pull Excitation and Continuous-Time $\Delta\Sigma$ T-D ReadoutarXiv · 2026
- Variability in Tissue Interface Temperature During Pulse Field Atrial Ablation: Implications for Real-Time Contact AssessmentCirculation Arrhythmia and Electrophysiology · 2026
- Complex-Polynomial Compensation of Wideband Harmonic Measurement Errors: A Closed-Loop Fluxgate Sensor Model StudyEnergies · 2026
- Recording sampling rate as a negative-control exposure for detecting measurement-driven bias in heart-rate-variability-outcome associations: a quantitative bias analysis across ICU waveform databasesmedRxiv · 2026