Advancing endodontic imaging: a high-resolution comparison between CBCT and micro-CT in the danger zone.

Gabriel de Toledo Telles-Araujo, Raimundo Sales de Oliveira Neto, Mariana Quirino Silveira Soares, Heitor Marques Honório, Viviane Almeida Sarmento, Marco Antônio Hungaro Duarte, Izabel Regina Fischer Rubira-Bullen

School of Medicine, Federal University of Bahia, Salvador, Brazil. · Department of Operative Dentistry, Endodontics, and Dental Materials, Bauru School of Dentistry, University of São Paulo, Bauru, Brazil.

Key points

From the paper's conclusion
  • CBCT systematically underestimates dentin thickness in the DZ and is not metrically equivalent to micro-CT for submillimetric measurements. Measurement accuracy is influenced by acquisition parameters and cannot be explained by voxel size alone.

Abstract

OBJECTIVE: This study evaluated the accuracy of cone-beam computed tomography (CBCT) protocols for measuring submillimetric dentin thickness in the danger zone (DZ) of mandibular molars, using micro-computed tomography (micro-CT) as the reference standard. MATERIALS AND METHODS: Fifty-one extracted mandibular molars were scanned using micro-CT and two CBCT devices under different acquisition protocols (Standard, High Fidelity, High Resolution [HIRE], and Imaging System [i-CAT]). After three-dimensional image registration, dentin thickness in the DZ, located 2 mm below the furcation, was measured in both mesial canals. RESULTS: All CBCT protocols significantly underestimated dentin thickness compared with micro-CT (p < 0.05), with discrepancies ranging from 4.5 to 8.6%. Underestimation was more pronounced in the mesiolingual canal. The HIRE protocol (0.125 mm voxel size) showed the smallest discrepancy, whereas the i-CAT protocol showed the largest. CONCLUSIONS: CBCT systematically underestimates dentin thickness in the DZ and is not metrically equivalent to micro-CT for submillimetric measurements. Measurement accuracy is influenced by acquisition parameters and cannot be explained by voxel size alone.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers