Potential Overestimation of Uveoscleral Outflow from Goldmann Tonometry Bias.
Sean McCafferty
University of Arizona College of Medicine, Tucson, AZ, USA.
Key points
From the paper's conclusion- Goldmann applanation tonometry (GAT) underestimates manometric IOP by about 4-5 mmHg.
Abstract
In living humans, uveoscleral outflow (Fu) is calculated, not measured. Except in rare tracer studies it is the residual of the modified Goldmann equation, Fu = F - C(IOP - Pev), which treats Fu as pressure-independent and assigns all pressure-dependent outflow to facility (C). Goldmann applanation tonometry (GAT) underestimates manometric IOP by about 4-5 mmHg. Because F, C, and Pev are measured independently of GAT, that low IOP input understates C(IOP - Pev) and overstates the residual reported as pressure-independent uveoscleral outflow. With representative values, GAT 15.5 mmHg yields Fu of about 50% of flow; manometric 20 mmHg yields about 9%, in line with human tracer data. A pressure-dependent uveoscleral component, if present, already sits inside C and cannot be separated by this method. The same bias has likely overstated the pressure-independent effect of prostaglandin analogs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- What Do Infrequent Goldmann Applanation Tonometry Measurements Reveal About Intraocular Pressure Profiles After Nonpenetrating Glaucoma Surgery?Translational vision science & technology · 2026
- Aqueous humor outflow facility in human eyes: History, techniques, and clinical implications.Survey of ophthalmology · 2026
- Clinical evaluation of corneal elasticity and true IOP using a dual-applanation tonometerFrontiers in Ophthalmology · 2026
- Effect of iStent inject on unmedicated intraocular pressure in glaucoma: exploratory analysis of clinical trial datamedRxiv · 2026
- Transient Corneal Response After Goldmann Three-Mirror Gonioscopy and Its Limited Effect on Optical Biometry and Calculated Intraocular Lens Power: A Prospective Within-Subject Controlled Observational Study.Medicina (Kaunas, Lithuania) · 2026
- Comparison of Intraocular Pressure Measurements Using iCare Rebound Tonometry and Goldmann Applanation Tonometry: A Systematic Review and Meta-Analysis.Journal of glaucoma · 2026