Highly Repeatable Measurement and Mechanical Modeling of Axial Insertion and Separation Forces in Hemodialysis Luer Connectors.
Tomohiro Matsui, Atsushi Sakurai
Department of Clinical Engineering, Faculty of Health Care Sciences, Jikei University of Health Care Sciences, Osaka, Japan. · Independent Researcher, Osaka, Japan.
Key points
From the paper's conclusion- Clinical disconnection risk was not directly quantified. Under controlled dry, axial loading, Luer connectors generally separated at lower forces than the preceding insertion force. Axial retention differed among connector combinations and was lower in check valve-equipped cannulas. Evaluating connection stability may require consideration of mechanical engagement characteristics alongside dimensional standards.
Abstract
PURPOSE: Unintended Luer connector disconnection during hemodialysis can cause blood loss and venous air embolism and persists despite Luer-lock mechanisms. The mechanical relationship between insertion force (Fin ) and separation force (Fsp ) remains unclear. PATIENTS AND METHODS: Fin and Fsp were measured under dry, axial loading using an automated linear stage and a six-axis force sensor, and a mechanical model based on Coulomb's law of friction was derived. Five male bloodline connector types and seven female needle or cannula connector types, including check valve-equipped cannulas, were tested in 35 combinations. Three female specimens per type were reused across male connectors, yielding 2001 cycles with insertion forces below 50 N. Linear mixed-effects models accounted for repeated measurements within specimens. RESULTS: Fsp increased approximately linearly with Fin across all combinations (per-combination slopes, 0.57-0.99). Without a check valve, the pooled slope was 0.855 (95% confidence interval, 0.817-0.894); however, Fsp exceeded Fin in 20% of cycles, primarily at low insertion forces. With a check valve, the relationship had a negative intercept (Fsp = 0.758 Fin - 4.07 N), and certain combinations showed unstable engagement even at insertion forces near 10 N. The model yielded Fsp /Fin = tan(ρ0 - θ)/tan(ρ' + θ), where ρ' is the insertion friction angle, ρ0 denotes the static friction angle during engagement, and θ is the Luer taper angle. This ratio is below 1 when friction remains constant. The observed slopes corresponded to static friction coefficients of approximately 0.1-0.4, comparable to reported values for plastics. CONCLUSION: Clinical disconnection risk was not directly quantified. Under controlled dry, axial loading, Luer connectors generally separated at lower forces than the preceding insertion force. Axial retention differed among connector combinations and was lower in check valve-equipped cannulas. Evaluating connection stability may require consideration of mechanical engagement characteristics alongside dimensional standards.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantifying User-Applied Forces on Prefilled Syringe Luer Cone: A Human Factors Ergonomics Study.PDA journal of pharmaceutical science and technology · 2026
- Blood Flow at Different Occlusion Pressures Using Multichambered Blood Flow Restriction Cuffs in the Arm and LegThe Journal of Strength and Conditioning Research · 2026
- Force measurement in circular external fixation frames reflects construct stability: A laboratory proof-of-concept study.Injury · 2026
- Preoperative Tranexamic Acid, Tourniquet Use, and Recorded Intraoperative Blood Loss in Total Knee ArthroplastySurgeries · 2026
- Bench Testing of an 8 N Slip Resistance Specification for the Handle Shaft Locking Interface of an Articulated Laparoscopic InstrumentmedRxiv · 2026
- Interchangeability Between Pleth Variability Index and Internal Jugular Vein Variability Index During Laparoscopic Living Donor Nephrectomy: A Prospective Observational Concordance StudymedRxiv · 2026