Medical devices (Auckland, N.Z.)

Multicenter Randomized Trial of Tactile Imaging versus Palpation for Neuraxial Insertion-Site Identification.

Dustin P LeBlanc, Robert B Wildman, Katherine H Hatter, Rohini D Samudralwar, Danielle L White, Jessica L Smith, George DeMuth, Yashar Ganjeh,

Department of Emergency Medicine, College of Medicine, Medical University of South Carolina, Charleston, SC, USA. · Department of Neurology, College of Medicine, Medical University of South Carolina, Charleston, SC, USA.

Randomised · n=90

Key points

From the paper's conclusion
  • Insertion attempts were numerically lower with device assignment in the ITT analysis and significantly lower with the device in the prespecified PP and exploratory AT analyses. The primary ITT superiority criterion was not met, and ITT procedure success was numerically lower with device assignment without a statistically significant difference. No serious or device-related adverse events were observed. These findings support tactile imaging as a potential bedside adjunct for neuraxial insertion-site identification.

Abstract

PURPOSE: To evaluate the performance of a tactile imaging device, VerTouch™ (IntuiTap Medical, Houston, TX, USA), compared with the standard palpation technique for insertion-site identification in neuraxial procedures. PATIENTS AND METHODS: In this multicenter, open-label, randomized superiority trial, adults undergoing neuraxial procedures across emergency medicine, neurology, and anesthesiology departments at three academic medical centers were randomized to insertion-site identification using a tactile imaging device or standard palpation; the intent-to-treat (ITT) analysis included 90 adults (42 device; 48 palpation). Proceduralists and site outcome observers were not blinded to treatment assignment. The primary outcome was the number of needle insertion attempts. RESULTS: In the primary ITT analysis, mean insertion attempts were 1.8 with device assignment and 2.2 with palpation (mean difference -0.4; 95% bootstrap confidence interval -1.2 to 0.3; one-sided p=0.133); the prespecified superiority criterion was not met. Insertion counts were significantly lower with the device in the prespecified per-protocol (PP) analysis (1.2 versus 2.1; one-sided p=0.005) and exploratory as-treated (AT) analysis (1.3 versus 2.2; two-sided p=0.003). These analyses were non-confirmatory. First-insertion success was higher with the device in all three analyses (ITT 69.4% versus 54.3%; PP 84.0% versus 61.0%; AT 78.6% versus 57.1%). Procedure success was 75.0% with device assignment and 89.1% with palpation in the ITT analysis (p=0.084); in the exploratory AT analysis, rates were 89.3% and 88.5%, respectively. No serious or device-related adverse events were observed. CONCLUSION: Insertion attempts were numerically lower with device assignment in the ITT analysis and significantly lower with the device in the prespecified PP and exploratory AT analyses. The primary ITT superiority criterion was not met, and ITT procedure success was numerically lower with device assignment without a statistically significant difference. No serious or device-related adverse events were observed. These findings support tactile imaging as a potential bedside adjunct for neuraxial insertion-site identification.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Multicenter Randomized Trial of Tactile Imaging versus Palpation for Neuraxial Insertion-Site Identification — Science Skim