Validation of fully automated myocardial T1 and synthetic hematocrit-based extracellular volume quantification on 3T cardiac MRI: a retrospective cross-sectional study.

Weerapat Kositanurit, Nonthikorn Theerasuwipakorn, Tanawin Suksiriworaboot, Sopida Thammongkolchai, Maethas Panyawongkhanti, Yongkasem Vorasettakarnkit, Chonthicha Lerdkhonsan, Monravee Tumkosit,

Department of Physiology, Faculty of Medicine, Chulalongkorn University, King Chulalongkorn Memorial Hospital, Bangkok, Thailand. · Division of Cardiovascular Medicine, Department of Medicine, Faculty of Medicine, Chulalongkorn University, Cardiac Center, King Chulalongkorn Memorial Hospital, Bangkok, Thailand.

Observational · n=450

Key points

From the paper's conclusion
  • Fully automated ECV quantification demonstrated strong correlation and agreement with the manual method within the normal Hct range. However, synthetic Hct-derived ECV was less reliable at abnormal Hct levels, particularly at Hct >45%. Further external validation and refinement of the synthetic Hct model are warranted to support broader clinical application.

Abstract

BACKGROUND: Automated post-processing software has been developed to quantify various cardiac magnetic resonance (CMR) parameters, including native T1 and extracellular volume (ECV) using synthetic hematocrit (Hct). However, validation studies of these automated measurements remain limited. We aimed to validate a fully automated workflow integrating myocardial T1 analysis and synthetic Hct for ECV quantification on 3T CMR, using manual ECV calculation based on venous Hct as the reference method. METHODS: This retrospective, cross-sectional study included 65 adults with low-to-intermediate risk of atherosclerotic cardiovascular disease who underwent stress 3T CMR. Manual ECV was calculated using manually contoured myocardial T1 values and venous Hct obtained within 7 days of CMR. Automated analysis used Myomics-T1 software (Phantomics, Korea) to automatically generate myocardial and blood-pool T1 values and contours, with synthetic Hct derived from a regression model developed in an independent cohort of 450 participants. Correlation and agreement were assessed using Pearson correlation, intraclass correlation coefficients (ICC), and Bland-Altman analysis. RESULTS: The mean age was 65.2±11.5 years, and 60% of participants were female. Automated myocardial native T1 and ECV showed excellent correlation with manual measurements (r=0.96 and 0.91, respectively) and excellent agreement (ICC =0.96 and 0.93, respectively). Automated myocardial native T1 was slightly higher than manual measurements (1,246±52 vs. 1,233±47 ms), whereas automated ECV was higher by 1.2 percentage points (29.0%±4.9% vs. 27.8%±4.6%). Synthetic Hct correlated with venous Hct (r=0.78; ICC =0.76) but demonstrated proportional bias. Agreement between synthetic Hct-derived and manual ECV was better within the normal Hct range (35-45%) but substantially reduced at Hct >45% (ICC =0.04). CONCLUSIONS: Fully automated ECV quantification demonstrated strong correlation and agreement with the manual method within the normal Hct range. However, synthetic Hct-derived ECV was less reliable at abnormal Hct levels, particularly at Hct >45%. Further external validation and refinement of the synthetic Hct model are warranted to support broader clinical application.