Evaluation and Classification of Atrial Arrhythmias from Signatures in Electrographic Flow Metrics.
Steven Castellano, Melissa H Kong, Joshua D'Arcy, Dmytro Perekrestenko, Kent R Nilsson, John D Hummel
Cortex, Inc., Menlo Park, CA, USA. · Department of Computer Science, Northwestern University, Chicago, IL, USA.
Key points
From the paper's conclusion- In conclusion, summative EGF metrics that characterize vector flow provide signature patterns of global atrial wavefront propagations in different atrial arrhythmias, which can be used to successfully determine the mechanisms of atrial arrhythmias.
Abstract
In patients undergoing atrial fibrillation (AF) ablation, it can be difficult at times to distinguish between AF, micro-re-entrant atrial flutters (AFLs), and focal atrial tachycardias (ATs). Understanding the mechanism of arrhythmogenesis can guide ablation lesion sets. Electrographic flow (EGF) mapping displays global atrial wavefront propagations in near-real time. Patterns of EGF can show both regions of stable flow direction over time and disorganized electrical wavefronts. In so doing, it can differentiate mechanisms of arrhythmogenesis. The objective of this study was to evaluate the EGF patterns and metrics that may be used to successfully differentiate mechanisms of arrhythmogenesis. A total of 2933 recordings from 151 patients who underwent EGF mapping were analyzed to determine metrics that could distinguish between mechanisms of arrhythmogenesis. These EGF-derived metrics include: (1) electrographic flow consistency (EGFC), (2) flow angle variability (FAV), (3) source activity, (4) atrial cycle length, (5) mean R-R correlation, and (6) disorganization score. The disorganization score varied the most between AF and AFL or AT (P < .0001), while the mean EGFC varied the most between AFL and AT (P < .0001). The mean FAV was found to be strongly correlated with both EGFC (r 2 = 0.767; P < .0001) and the disorganization score (r 2 = 0.552; P < .0001); therefore, the mean FAV was shown to contain the composite information of the other two metrics. A multilayer perceptron trained and tested on an equal number of recordings in each arrhythmia was also able to accurately determine the rhythm from previously unseen samples with 81% accuracy when given only the five remaining metrics. Additionally, it was 100% successful at distinguishing AF from the more organized rhythms. In conclusion, summative EGF metrics that characterize vector flow provide signature patterns of global atrial wavefront propagations in different atrial arrhythmias, which can be used to successfully determine the mechanisms of atrial arrhythmias.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Characterization of AF Recurrence and Substrate Health by Electrographic Flow (EGF) Consistency in EGF Mapping.Journal of cardiovascular electrophysiology · 2026
- Machine learning for enhancing the ablation outcomes using recurrence quantification analysis in human persistent atrial fibrillation.Frontiers in physiology · 2026
- Focal Activation Patterns in Atrial Fibrillation: Are We Seeing the Same Across Mapping Methods?Heart rhythm · 2026
- Novel predictors of late recurrence after catheter ablation for atrial fibrillation: from biomarkers to artificial intelligence models.Frontiers in cardiovascular medicine · 2026
- ECG-XPLAIM: eXPlainable Locally-adaptive Artificial Intelligence Model for arrhythmia detection from large-scale electrocardiogram dataFrontiers in Cardiovascular Medicine · 2025
- Generative recurrence analysis and metric learning for detection of atrial fibrillation from short ECG data.Chaos (Woodbury, N.Y.) · 2026