科研速览 · Science Skim继续刷下去 · Keep skimming →
◆ AAPS PharmSciTech2026-09-23

High-Throughput eSWIR Hyperspectral Imaging for Pharmaceutical Classification at Industrial Line Speeds.

Siamak Kashi, Yongjin Sung

原始摘要(英文原文)· Original abstract
Real-time quality control of pharmaceutical dosage forms requires rapid, non-destructive analytical technologies that provide chemical specificity while maintaining high throughput. Here, an extended short-wave infrared (eSWIR, 1000-2500 nm) line-scan hyperspectral imaging system (L-EOS 2.5e, Photon Etc.) was evaluated for spatially resolved classification of pharmaceutical tablets. The system was evaluated at integration times of 2, 0.2, and 0.02 ms using Standard Normal Variate normalization, Savitzky-Golay second-derivative preprocessing, and an error-correcting output codes support vector machine classifier. Product-level classification of 11 pharmaceutical products achieved accuracies of 99.99%, 100%, and 97.6% at 2, 0.2, and 0.02 ms, respectively. A more challenging manufacturer-level discrimination task involving three 500 mg acetaminophen products containing the same active pharmaceutical ingredient and labeled dose achieved accuracies of 99.7%, 98.6%, and 89.3% at the corresponding integration times. These results demonstrate the potential of eSWIR hyperspectral imaging for rapid, spatially resolved pharmaceutical product discrimination and high-throughput inspection.
读原文 · Read the paper ↗

AI 追问PRO

登录后使用 AI 追问

讨论区

登录后参与讨论

相关论文 · Related

High-Throughput eSWIR Hyperspectral Imaging for Pharmaceutical Classification at Industrial Line Speeds. — 科研速览 Science Skim