Effect of Ultra-Conformable Coatings on CMUT Behaviour in Biological Environments

Omar Najib Aboul-Hosn, Benedikt Szabo, Rob van Schaijk, Thomas Stieglitz

Key points

From the paper's conclusion
  • The thin PDMS layer therefore provides corrosion protection, preserves membrane dynamics, and offers acoustic‑impedance matching (≈ 1.5 MRayl), supporting its use as a low‑cost encapsulant for CMUT‑based wireless power transfer in liquid environments.

Abstract

Capacitive‑micromachined ultrasonic transducers (CMUTs) are promising wireless‑power sources for deep‑seated active implanted medical devices; yet their membranes are not made for implantation, and therefore require the application of a biocompatible coating to protect the patient as well as the membranes. This work investigates a thin (≈ 65 μm), medical‑grade silicone (PDMS) encapsulation as a simple, biocompatible protection layer. Three PCB‑mounted CMUT batches were examined: (i) uncoated (A1), (ii) coated with MED‑1000 (B1), and (iii) coated with MED2‑4213 (C1). Digital holographic microscopy (DHM) quantified membrane displacement in air (dry) and in 5 mL phosphate‑buffered saline (wet). In PBS the uncoated CMUT showed severe corrosion and a mean displacement of 1.05 ± 0.96 nm, whereas both silicone‑coated devices retained a mean displacement of ≈ 2.5 nm (≈ 44 % of the dry value) with no visual damage. A brief hydrophone test confirmed that the coating does not attenuate the dominant ~10 MHz acoustic component. The thin PDMS layer therefore provides corrosion protection, preserves membrane dynamics, and offers acoustic‑impedance matching (≈ 1.5 MRayl), supporting its use as a low‑cost encapsulant for CMUT‑based wireless power transfer in liquid environments.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers