Validation of an LC-Q-TOF "HiRes Mass-fix" assay for improved sensitivity and specificity of M-protein detection.
Ryan W Pearce, Mindy C Kohlhagen, Mark A Martinez, Maria A V Willrich, Dragan Jevremovic, Jonathan Coker, David L Murray
Division of Clinical Biochemistry and Immunology, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN, USA.
Key points
From the paper's conclusion- HiRes Mass-Fix was successfully validated and demonstrated to increase M-protein detection in samples negative by Mass-Fix. IgG recycling may explain the blood-positive/BM-negative discordance, and serial measurements will be needed to distinguish residual M-protein from ongoing production. Blood-based M-protein measurement is a less invasive, lower-cost alternative to bone marrow biopsy for ongoing patient monitoring.
Abstract
OBJECTIVES: To validate a blood-based liquid chromatography quadrupole time-of-flight mass spectrometry (MS) method "HiRes Mass-Fix" as a fit-for-purpose, patient-specific method for detecting low level M-proteins in patients who have reached negativity by MALDI-TOF MS "Mass-Fix". METHODS: Analytical validation included precision, accuracy, analytical measurement interval, limit of detection, specificity, ion suppression, carryover, and extracted-sample stability. Clinical performance was assessed by comparing 174 patients with bone marrow (BM) measurable residual disease (MRD) by next-generation flow cytometry (NGF) to both Mass-Fix and HiRes Mass-Fix. RESULTS: HiRes Mass-Fix demonstrated acceptable precision, accuracy, linearity, specificity, and stability for its intended use, with an average limit of detection of 3.8 mg/L. HiRes Mass-Fix detected M-proteins in 41 of 58 Mass-Fix negative samples. This raised blood MRD positivity from 62 to 89 % against a BM NGF positivity rate of 51 %. Blood-BM agreement fell from 71 % (kappa 0.43) by Mass-Fix to 58 % (kappa 0.15) by HiRes Mass-Fix, driven by blood-positive/BM-negative discordance. IgG was overrepresented among discordant cases (68 vs. 45 %, p=0.003). CONCLUSIONS: HiRes Mass-Fix was successfully validated and demonstrated to increase M-protein detection in samples negative by Mass-Fix. IgG recycling may explain the blood-positive/BM-negative discordance, and serial measurements will be needed to distinguish residual M-protein from ongoing production. Blood-based M-protein measurement is a less invasive, lower-cost alternative to bone marrow biopsy for ongoing patient monitoring.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- B-350 Clinical Validation of High-Resolution Mass Spectrometry for Blood-Based Minimal Residual Disease Detection to Reduce the Need for Bone Marrow AspirationClinical Chemistry · 2026
- Current Evidence for Integrating Mass Spectrometry into Clinical M-Protein Assessment in Plasma Cell Dyscrasias.Clinical chemistry · 2026
- Blood-based MALDI-TOF mass spectrometry versus bone marrow MRD for monitoring multiple myeloma: CASSIOPEIA study results.Blood · 2026
- Blood-Based Mass Spectrometry in Plasma Cell Disorders: From Residual Secretory Activity to Multimodal Residual Disease Assessment.Critical reviews in oncology/hematology · 2026
- Automated peak detection, isotyping and quantitation of M-proteins by MALDI-TOF MS for replacement of immunofixation and serum protein electrophoresisJournal of Immunological Methods · 2026
- One clone, one mass: The rise of mass spectrometry in the diagnosis and monitoring of monoclonal gammopathies.Clinica chimica acta; international journal of clinical chemistry · 2026