A generalizable fit-for-purpose workflow for multiplex targeted mass-spectrometry of protein-based blood biomarkers: validation and clinical application to neurodegenerative diseases in a 400-patient cohort.
Florine Leipp, Jérôme Vialaret, Ann-Christin Niehoff, Marie Duchiron, Karim Bennys, Sylvain Lehmann, Christophe Hirtz
Shimadzu France, Noisiel, France. · IRMB-PPC, INM, CHU Montpellier, INSERM, University of Montpellier, Montpellier, France.
Key points
From the paper's conclusion- This LC-MRM workflow combines analytical rigor with streamlined processing for large-scale multiplex protein biomarker quantification in plasma, with demonstrated translational potential for broader clinical research application.
Abstract
OBJECTIVES: Accurate quantification of endogenous plasma proteins remains a major challenge for clinical translation of proteomic biomarkers, particularly in neurodegenerative diseases where low-abundance targets coexist with highly abundant proteins in a complex matrix. Fit-for-purpose validation frameworks compatible with ICH M10 guidelines offer an adapted strategy for multiplex targeted mass-spectrometry assays. We developed and validated a streamlined LC-MRM workflow for plasma biomarker quantification and evaluated its analytical applicability and biological coherence in a deeply phenotyped Alzheimer's disease (AD) cohort. METHODS: Twenty-four proteotypic peptides representing 20 protein biomarkers were selected based on prior evidence and mass spectrometry (MS) detectability. Cleavable, winged stable isotope-labeled (SIL) peptides were used as surrogate analytes and internal standards to account for digestion and downstream analytical variability in the absence of a true blank plasma matrix. Analytical validation assessed linearity, accuracy expressed as percentage of nominal concentration, bias, precision, carry-over, matrix effects, digestion efficiency, stability, and inter-operator reproducibility. The panel was applied to 400+ samples from the ALZAN cohort (NCT05427448), classified by cerebrospinal fluid amyloid/tau/neurodegeneration (AT[N]) status. RESULTS: Twenty-one of 24 peptides met acceptance criteria; 17 were endogenously measurable in plasma. Linearity (R2≥0.98), precision (CV≤20 %), and accuracy (80-120 % of nominal concentration) were consistently achieved. Matrix effects were substantial but were compensated by the cleavable SIL strategy. PARK7 and Clusterin were significantly decreased in AD vs. controls (p<0.01), and Transthyretin differed between non-AD neurodegenerative conditions and controls. AT-stratified analyses identified differences in PARK7, Clusterin, ApoE, and Osteopontin across groups. CONCLUSIONS: This LC-MRM workflow combines analytical rigor with streamlined processing for large-scale multiplex protein biomarker quantification in plasma, with demonstrated translational potential for broader clinical research application.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Multicenter evaluation of label-free quantification in human plasma on a high dynamic range benchmark setNature Communications · 2025
- Development of Highly Multiplex Targeted Proteomics Assays in Biofluids Using a Nominal Mass Ion Trap Mass SpectrometerMolecular & Cellular Proteomics · 2026
- Analytical validation and clinical performance of a chemiluminescence immunoassay panel for plasma biomarkers in Alzheimer's disease: a preliminary study.Clinica chimica acta; international journal of clinical chemistry · 2026
- Short-interval variability and test-retest reliability of Alzheimer's disease plasma biomarkers in a clinically heterogenous cohort.Alzheimer's research & therapy · 2026
- Silent clues in blood: Proteomic approach for Alzheimer's disease.International review of neurobiology · 2026
- Analytical Methods for Fluid Biomarkers in Alzheimer’s Disease from Discovery to Clinical ImplementationInternational Journal of Molecular Sciences · 2026