Pre-analytical and analytical validation of calcitonin measurement in fine needle aspiration cytology washout.
Amber Mater, Annemieke C Heijboer, Anita Boelen, Pim de Graaf, Koen M A Dreijerink, Jacquelien J Hillebrand
Department of Laboratory Medicine, Endocrine Laboratory, Amsterdam UMC location Vrije Universiteit Amsterdam, Amsterdam, The Netherlands. · Amsterdam Gastroenterology, Endocrinology and Metabolism, Amsterdam, The Netherlands.
Key points
From the paper's conclusion- This study showed that MultiDiluent was the optimal solvent to use for calcitonin measurements in a non-serum matrix using the Atellica calcitonin immunoassay. CytoLyt performed suboptimal, while the performance of NaCl was poor. Further research is needed to validate the recommended work-up using patient-derived material.
Abstract
OBJECTIVES: Current guidelines advise calcitonin measurements in washout from fine needle aspiration cytology (FNAC) upon inconclusive cytological examination of medullary thyroid cancer (MTC). (Pre-)analytical validation studies of calcitonin measurements in FNAC washout are however missing. In this study, we examined (pre-)analytical aspects of calcitonin measurements using three washout solvents, aiming to provide insights for implementation of calcitonin measurements in FNAC washout as a supplementary test for diagnosing and monitoring patients with MTC. METHODS: Validation studies were performed using CytoLyt, NaCl and MultiDiluent as washout solvents. Precision (intra- and inter-assay coefficients of variation (CV)), accuracy (linearity and recovery) and storage stability of each solvent were assessed after spiking with calcitonin standard. Calcitonin measurements were performed using the serum CALCT immunoassay (Siemens, Atellica). Decision criteria were set based on CLSI and manufacturer's manual. RESULTS: Measurements of calcitonin in MultiDiluent met all decision criteria: intra- and inter-assay CVs were ≤5.5 % and ≤9 %, serial dilutions were linear (R2=0.997) and median recovery was 148 % with limited variation. Furthermore, calcitonin concentrations remained stable (<10 % change) after storage. Measurements in CytoLyt were precise and accurate, but unstable upon storage. Measurements in NaCl only met the criteria for intra-assay CV and linearity. CONCLUSIONS: This study showed that MultiDiluent was the optimal solvent to use for calcitonin measurements in a non-serum matrix using the Atellica calcitonin immunoassay. CytoLyt performed suboptimal, while the performance of NaCl was poor. Further research is needed to validate the recommended work-up using patient-derived material.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Diagnostic Accuracy of Calcitonin, Procalcitonin, proGRP, miR-375 and FNA-Washout Calcitonin for Medullary Thyroid Carcinoma: A Systematic Review and Diagnostic Meta-Analysis.Life (Basel, Switzerland) · 2026
- Ultrasound-guided fine-needle aspiration cytology combined with washout fluid thyroglobulin measurement in the diagnosis of cervical lymph node metastasis in papillary thyroid carcinoma: research advances.Frontiers in immunology · 2026
- Is Calcitonin at 6–12 Months an Appropriate Criterion to Define Cure in Medullary Thyroid Carcinoma?World Journal of Surgery · 2026
- Longitudinal Changes in Urinary ExoQuick-TC-Enriched Fraction-Derived Tryptic Peptides Following Thyroidectomy in Well-Differentiated Thyroid Cancer: The t-UXEDO Study.Life (Basel, Switzerland) · 2026
- 2026 ETA Consensus Statement on the Interpretation and Management of Calcitonin Levels in Patients with Thyroid Nodules.European thyroid journal · 2026
- Multicenter evaluation of analytical performance and onboard stability of Mindray 10 concentrated biochemical reagentsJournal of Laboratory Medicine · 2026