Persistent residual ellipticity in waveplate-compensated polarization-resolved SHG microscopy on a commercial multiphoton microscope.
David Nguyen, Jeffrey P Wilde, Virginie Uhlmann, Daniel James Smith, Justine Kusch-Wieser, Valeria Zanrè, Jakob Schwiedrzik, Gábor Csúcs
Scientific Center for Optical and Electron Microscopy (ScopeM), ETH Zürich, Zürich, Switzerland. · Department of Electrical Engineering, Stanford University, Stanford, California, United States of America.
Key points
From the paper's conclusion- Here, we reproduce an established automated waveplate compensation method, based on a rotating half-waveplate combined with a compensating quarter-waveplate, on a commercial Leica TCS SP8 MP multiphoton microscope, which is the same type of instrument the method was originally developed on.
Abstract
Polarization-resolved second harmonic generation microscopy provides structural information about non-centrosymmetric biological samples such as collagen. It involves illuminating the sample with a focused laser beam having a variable linear polarization angle and recording the second harmonic signal as a function of this angle. However, accurate linear polarization control is challenging due to ellipticity introduced by reflections from mirrors and dichroic mirrors in the optical path. Waveplate-based compensation has emerged as the standard approach to address these distortions, but its effectiveness for quantitative measurements remains incompletely characterized. Here, we reproduce an established automated waveplate compensation method, based on a rotating half-waveplate combined with a compensating quarter-waveplate, on a commercial Leica TCS SP8 MP multiphoton microscope, which is the same type of instrument the method was originally developed on. Despite significant effort, we consistently observed substantial unwanted residual polarization ellipticity, with amplitudes up to 0.24, persisting under optimal waveplate configurations. Our simulation analysis suggests that this limitation may arise from wavelength-dependent dichroic mirror birefringence combined with the broad spectral bandwidth (10 nm to 20 nm full width at half maximum) of femtosecond laser pulses. While the approach investigated here can compensate a single wavelength, different spectral components within the pulse experience different phase retardations from wavelength-dependent optical elements, potentially resulting in residual ellipticity that cannot be eliminated. Our simulations qualitatively reproduced key features of the experimental observations. These findings have important implications for quantitative polarization-resolved second harmonic generation microscopy and suggest that alternative approaches, including specimen rotation or picosecond laser sources with narrower bandwidth, should be investigated for applications requiring precise polarization control. To facilitate community investigation of these effects, we provide open-source analysis code and simulation files.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Wavelength and polarization optimization through Stokes parameters for second harmonic generation microscopy in corneal tissue.Optics letters · 2026
- Multimodal video-rate polarization and phase microscopy by asymmetric spectrally structured illumination.Biomedical optics express · 2026
- Polarized Light Microscopy for Quantitative Assessment of Sarcomere-Scale Structural Order in Cardiac Tissue.Journal of biophotonics · 2026
- Two-Photon Fluorescence and Second-Harmonic Generation Microscopy: Principles and Biomedical Applications.Journal of biophotonics · 2026
- Correlation-resolved sum-frequency generation microscopy of biological tissue with bright twin beamsbioRxiv · 2026
- Unambiguous polarimetric reconstruction in single-input polarization-sensitive OCT via spectral polarization spreading.Biomedical optics express · 2026