Digital measurement of ocular microtremor in Parkinson’s Disease: Analytical and clinical validation
Lisa Graham, Rodrigo Vitório, Patrick Tait, Richard Walker, Alan Godfrey, Rosie Morris, Samuel Stuart
Northumbria University · Northumbria Healthcare NHS Foundation Trust
Key points
From the paper's conclusion- This pilot study aimed to examine the analytical and clinical validation of OMT measurement via the iTremor ONE in people with Parkinson's Disease (PwPD).
Abstract
Ocular microtremor (OMT) is an involuntary fixational eye movement linked to brainstem activity. OMT is thought to have a mean frequency range of 70-90 Hz in healthy adults. Previous research suggests OMT may be reduced in neurological diseases like Parkinson's Disease. Historically, OMT has been measured invasively in specialist laboratories using lengthy and expensive protocols. Developments now allow for OMT measurement quickly and non-invasively using hand-held technology (i.e., iTremor ONE). This pilot study aimed to examine the analytical and clinical validation of OMT measurement via the iTremor ONE in people with Parkinson's Disease (PwPD). 33 PwPD and 31 age matched healthy controls participated in this study. For analytical validation, 22 PwPD completed a test re-test reliability assessment of OMT measurement, assessed using interclass correlation coefficients (ICC). For clinical validation, OMT frequency in PwPD (n = 33) was compared to controls. Correlations were explored with demographics and clinical scales. Additionally, 24 PwPD were tested 'OFF' (12hr withdrawal) and 'ON' their anti-Parkinson's (dopaminergic) medication to compare OMT response to a known intervention. The iTremor ONE demonstrated excellent test-retest reliability (ICC > 0.9) for measuring OMT frequency in PwPD. Mean OMT frequency was significantly lower in PwPD (63.78 ± 4.82 Hz) compared to controls (69.44 ± 6.47 Hz, p < .001), with good discriminative ability (AUC 0.75-0.77). OMT frequency correlated with age in both groups and with specific motor features (speech, facial expression, gait) in PwPD. No significant differences in OMT frequency were observed between 'OFF' and 'ON' dopaminergic medication states. This is the first study to demonstrate that a non-invasive hand-held device can reliably measure OMT in PwPD and presents OMT analytical and clinical validation evidence. OMT frequency may provide a supporting measure for diagnosis or screening. Further research is required to understand the neural mechanisms underpinning OMT in PwPD and the role it could play in clinical practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Spontaneous eye behavior is an objective marker of parkinsonian severity.bioRxiv : the preprint server for biology · 2026
- Multimodal eye-tracking reveals impaired inhibitory control and reduced preparatory pupil dilation in essential tremor.Brain communications · 2026
- Reliability of Myotonometric Measurements and Their Associations with Motor Performance in Parkinson's Disease.Sensors (Basel, Switzerland) · 2026
- Eye-Tracking Assessment of Visual, Oculomotor, and Pupillary Biomarkers During Reading in Parkinson's Disease.Life (Basel, Switzerland) · 2026
- Less is more: comparison of automated spiral-drawing metrics with clinician-rated tremor scores after MRgFUS thalamotomy.Frontiers in neurology · 2026
- A Stand-Alone Wearable Measurement System for Continuous Hand Tremor Monitoring.Sensors (Basel, Switzerland) · 2026