From the paper's conclusion- The results demonstrate spatially adaptive compensation by integrating calibration-derived physical priors into a computational reconstruction framework compatible with the existing imaging system.
Abstract
Field-dependent blur and frequency-response loss limit miniature confocal endoscopic imaging. We introduce a physics-prior-guided dual-branch residual network that combines the degraded image with normalized field coordinates and calibration-derived MTF, PSF, SNR, and intensity-response maps. For the representative peripheral resolution-target region, the peak-to-valley difference increased from 0.23 to 0.36, and image-derived MTF50 increased from 16.7 to 40.4 lp/mm. Across 196 patches extracted from the four displayed source images, the proposed method achieved an NRMSE of 0.1394 and an SSIM of 0.8264. The results demonstrate spatially adaptive compensation by integrating calibration-derived physical priors into a computational reconstruction framework compatible with the existing imaging system.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Blind spatially varying PSF estimation for fluorescence microscopyarXiv · 2026
- Adaptive aberration-corrected quantitative phase microscopy via physics-informed deep learningAdvanced Photonics · 2025
- Coordinate-conditioned Deconvolution for Scalable Spatially Varying High-Throughput Imaging.IEEE transactions on computational imaging · 2026
- High‐Fidelity, Adaptive Label‐Free Imaging via Color‐Multiplexing With Analytical Aberration EstimationLaser & Photonics Reviews · 2026
- EID-Net: Endoscopic Illumination-aware Degradation Correction NetworkarXiv · 2026
- Differentiable Inverse Rendering Recovers Intrinsic Properties from Intravascular Optical Coherence TomographyarXiv · 2026