Optics express

Compact laser power measurement at tens of watts via amplified radiation pressure.

Atlas Rhode, Benjamin Nienhouse, Brian Simonds, Paul Williams, Kyle Rogers

National Institute of Standards and Technology, 325 Broadway, Boulder, CO 80305, USA.

Key points

From the paper's conclusion
  • An amplified radiation-pressure power measurement was conducted using a compact, high-amplification laser-pressure optic (HALO-c), with an expanded uncertainty of 0.93% at a power of 100 W (95% confidence interval).

Abstract

An amplified radiation-pressure power measurement was conducted using a compact, high-amplification laser-pressure optic (HALO-c), with an expanded uncertainty of 0.93% at a power of 100 W (95% confidence interval). A commercial force balance was used to measure the photon pressure amplified by 10 reflections for powers ranging from 5 W - 200 W. These low powers were measurable due to the high amplification factor of the light being measured and the low noise on the force balance. The HALO-c was constructed with all but two components being off-the-shelf. Our measurements were validated through comparison with a calibrated thermopile transfer standard; we found agreement within the measurement uncertainty at powers down to 5 W.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers