Biomedical optics express

Glass wedge strabismic imaging for scanning instruments.

Gastón A Ayubi, Karteek Kunala, Yuning Xia, Alfredo Dubra

Byers Eye Institute, Stanford University, Palo Alto, CA 94303, USA.

Key points

From the paper's conclusion
  • We propose using a custom glass wedge in reflection to generate two illumination beams for implementing strabismic imaging in point- and line-scanning imaging instruments.

Abstract

We propose using a custom glass wedge in reflection to generate two illumination beams for implementing strabismic imaging in point- and line-scanning imaging instruments. We present wedge design formulae that ensure beam overlap at pupil conjugates while achieving a desired beam separation at image conjugates. These formulae provide, for an in-plane wedge rotation, the rotation angle of the primary reflection from the second wedge surface about the primary reflection from the first surface, and the resulting small residual separation at pupil conjugates. We demonstrate the approach in an adaptive optics scanning light ophthalmoscope with reflectance confocal and non-confocal imaging channels. This low-cost, low-complexity method for generating strabismic illumination from a single collimated beam provides long-term mechanical stability and is broadly applicable across imaging modalities.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Glass wedge strabismic imaging for scanning instruments — Science Skim