Biomedical optics express

Optical phantom for resolution evaluation of non-contact corneal microscopy techniques.

Kenneth Marcelino, Jason Datta, Momoka Sugimura, Carmella Ocaya, Rafael Romero, Yongjun Kim, Mac Rodriguez, Milind Rajadhyaksha,

James C. Wyant College of Optical Sciences, University of Arizona, Tucson, Arizona 85721, USA. · Department of Biomedical Engineering, University of Arizona, Tucson, Arizona 85721, USA.

Key points

From the paper's conclusion
  • We present the development of a corneal phantom designed to enable resolution assessment for non-contact corneal microscopy.

Abstract

We present the development of a corneal phantom designed to enable resolution assessment for non-contact corneal microscopy. The phantom was fabricated by casting a UV-curable epoxy mixed with 400-nm-diameter gold nanoparticles into a polydimethylsiloxane (PDMS) mold with a curved surface to achieve a cornea-like geometry. The radius of curvature of the final manufactured phantom was 8.14-8.17 mm with a refractive index of 1.39, similar to the adult human cornea. The optical phantom was used to evaluate the resolution of a prototype non-contact confocal microscope, the portable in vivo confocal ophthalmoscope (PICO). PICO images of the optical phantom allowed for evaluation of the resolution change as a function of depth.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Optical phantom for resolution evaluation of non-contact corneal microscopy techniques — Science Skim