Applied optics

Improving probe-beam stability in spherical-cell atomic magnetometers using 4f spatial filtering.

XueJing Liu, ZhenXiang Tang, Yang Li

The Key Laboratory of Optical Technology and Instrument for Medicine, Ministry of Education, School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China. · The School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China.

Key points

From the paper's conclusion
  • We propose a pre-positioned, 4f spatial-filtering system to suppress high-angle components of the probe beam and improve its stability against relative position disturbances between the beam and the vapor cell.

Abstract

Spherical vapor cells introduce position-dependent refraction and variations in the incident-angle distribution that can destabilize the probe-beam polarization readout in atomic magnetometers. We propose a pre-positioned, 4f spatial-filtering system to suppress high-angle components of the probe beam and improve its stability against relative position disturbances between the beam and the vapor cell. A Jones-matrix model is developed to describe the effect of incident-angle variations on polarization, and experiments are performed to evaluate the extinction ratio and signal-to-noise ratio (SNR). With 4f filtering, the extinction-ratio variation is reduced by 15.67%-65.39%, while the SNR variation decreases by 19.21%-74.54% under vapor-cell position perturbations. Zemax simulations further confirm improved received-intensity stability.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers