Tetsu Nakaichi, Hiroyuki Okamoto, Mitsuhiro Kon, Kazuki Takaso, Ako Aikawa, Satoshi Nakamura, Kotaro Iijima, Takahito Chiba, Hiroki Nakayama, Tomonori Goka, Hiroshi Igaki
This preclinical phantom study demonstrates the feasibility of wide-area CCD-EPID transit-dose estimation for conventional long-SSD TBI. Agreement in homogeneous phantoms supports the calibration approach, while the positive RANDO bias identifies the need for same-day sensitivity normalization and further validation in repeated anthropomorphic, heterogeneous, lung-shielded, and full-clinical-field conditions before patient application.
BACKGROUND: Conventional extended-distance total body irradiation (TBI) is commonly verified using point dosimeters, which provide limited spatial information and require additional handling during a prolonged treatment procedure. A portable charge-coupled device (CCD)-based electronic portal imaging device (EPID) positioned downstream of the patient may provide wider-area transit information in a geometry in which a gantry-mounted EPID cannot readily be used.
PURPOSE: To develop and perform preclinical phantom validation of a CCD-EPID method for transit dose estimation during fixed-field long-SSD TBI and to identify the calibration and normalization factors relevant to subsequent patient validation.
METHODS: A portable CCD-based EPID was positioned at 400 cm, with water-equivalent or anthropomorphic phantoms centered at 350 cm. Direct-irradiation characteristics were evaluated for 4- and 10-MV photon beams, including signal response to dose-rate/delivered MU, field size, short-term frame reproducibility, and in-plane response. For ETD calibration, 6-36-cm-thick water-equivalent phantoms were irradiated with 10-MV photons using 1000 MU at 600 MU/min and a 10 × 10 cm2 jaw setting at isocenter, corresponding to approximately 35 × 35 cm2 at the phantom and 40 × 40 cm2 at the EPID. Summed EPID signals were related to midplane absorbed dose measured using a Farmer-type ionization chamber. ETD was independently compared with radiophotoluminescent dosimeters (RPLDs) in 12-, 24-, and 36-cm slab phantoms and at 11 locations in a RANDO phantom. Four separate RANDO irradiations were centered on the head, chest, abdomen, and pelvis.
RESULTS: The 10-MV EPID signal was highly linear with dose rate/delivered MU (r = .999) and field size (r = .995). Flatness was 1.4% and 1.3%, and symmetry was .3% and 1.4%, in the ceiling-floor and gun-target directions, respectively. The summed EPID signal-to-ionization-chamber dose relationship showed R2 = .999. ETD differed from RPLD dose by -1.6%, -1.4%, and .0% in the 12-, 24-, and 36-cm slab phantoms, respectively, within the expanded RPLD uncertainty. In the RANDO phantom, ETD was consistently higher than RPLD dose, with a mean relative difference of 7.7 ± 3.9%. Because calibration and RANDO verification used the same field size and the detector showed high dose/MU linearity, a field-size mismatch or gross signal nonlinearity is unlikely to be the dominant explanation. Heterogeneous attenuation and scatter, local path-length/ROI correspondence, and session-dependent CCD sensitivity are plausible contributors.
CONCLUSIONS: This preclinical phantom study demonstrates the feasibility of wide-area CCD-EPID transit-dose estimation for conventional long-SSD TBI. Agreement in homogeneous phantoms supports the calibration approach, while the positive RANDO bias identifies the need for same-day sensitivity normalization and further validation in repeated anthropomorphic, heterogeneous, lung-shielded, and full-clinical-field conditions before patient application.