Moving-average processing enables accurate quantification of time delay and compares the trending ability of cardiac output monitors with different response times
Yoshihiro Sugo, Ryoichi Ochiai
Nihon Kohden (Japan) · Toho University
Key points
From the paper's conclusion- Moving average processing over 20 to 30 min significantly enhanced concordance between esCCI and reference CCI, with APCI demonstrating similarly high concordance in the same time window.
Abstract
BACKGROUND: Continuous cardiac output (CCO) monitoring using pulmonary artery (PA) thermodilution and newly introduced beat-to-beat cardiac output (CO) monitoring technologies exhibits different response time delays. These differences can hinder accurate comparisons of their trending abilities. To address this, we applied moving average processing to the beat-to-beat CO monitor data to evaluate its effect on trending assessment accuracy. This study aimed to confirm the effectiveness of moving average processing for such comparisons. RESULTS: This was a single-center, retrospective, observational study conducted at a 916-bed university hospital. A total of 20 patients undergoing kidney transplantation were included. We analyzed the trending ability of arterial pressure cardiac index (APCI) and estimated continuous cardiac index (esCCI) relative to continuous cardiac index (CCI) derived from PA thermodilution. Trending ability was assessed using a Polar plot and Bland-Altman analyses. A wide range of moving average windows (0–60 min) was applied to APCI and esCCI. The polar concordance rate at 30° exceeded 92% for moving average windows between 20 and 30 min, with APCI peaking between 21 and 27 min. These improvements reflected both time-shifting and filtering effects of the moving average process. CONCLUSIONS: Moving average processing over 20 to 30 min significantly enhanced concordance between esCCI and reference CCI, with APCI demonstrating similarly high concordance in the same time window. This approach effectively compensates for differences in response time delays between CO monitoring modalities, enabling more accurate assessment of trending ability.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The map and the territory: a critical appraisal of pulse contour cardiac output monitoring.British journal of anaesthesia · 2026
- The effect of preload on Doppler systolic time in a large emergency department cohort with suspected sepsis.Intensive care medicine experimental · 2026
- Electrical Cardiometry for Non-Invasive Cardiac Output Monitoring in Neonatal and Pediatric Critical Care: A Bayesian Systematic Review and Meta-Analysis.Journal of intensive care medicine · 2026
- Detecting if change in end-tidal carbon dioxide signifies change in cardiac output for the prediction of fluid responsiveness in patients undergoing major abdominal oncosurgery: An observational study.Indian journal of anaesthesia · 2026
- Diagnostic Performance of Corrected Carotid Flow Time for Assessing Fluid Responsiveness in Patients With Septic or Hypovolemic Shock: A Prospective Diagnostic Accuracy Study.Cureus · 2026
- Iterative delay correction improves breath-hold cerebrovascular reactivity mapping in clinical populations.Imaging neuroscience (Cambridge, Mass.) · 2026