Clinical evaluation of a novel artificial intelligence-based image processing algorithm for two-dimensional angiography.
David C Madoff, Hector Tamez Aguilar, Michael J Angtuaco, Jeremy D Asnes, John D Barr, Donald F Frei, Jon C George, Eric J Hohenwalter,
Department of Radiology and Biomedical Imaging, Section of Interventional Radiology, Yale School of Medicine, New Haven, CT, USA. david.madoff@yale.edu. · Piedmont Heart Institute, Atlanta, GA, USA.
Key points
From the paper's conclusion- In this multicenter blinded reader study, AI-based image processing demonstrated non-inferior overall image quality compared with its predecessor across all evaluated imaging modes and patient subgroups, with superior performance for multiple image quality metrics in fluoroscopy, roadmap, overlay reference, and native acquisition imaging. These findings support the clinical utility of AI-based image processing for endovascular imaging and provide a foundation for future studies evaluating its impact on radiation dose and procedural outcomes.
Abstract
BACKGROUND: To evaluate the clinical performance of a novel artificial intelligence (AI)-based image processing algorithm for two-dimensional angiographic imaging compared with its predecessor across multiple imaging modes, anatomic regions, and endovascular specialties. MATERIALS AND METHODS: This retrospective, comparative, randomized, paired, and blinded multicenter reader study included anonymized images acquired during clinically indicated endovascular interventions at 11 institutions in the United States and Europe. A total of 553 image series from 182 patients representing 51 procedure types across interventional radiology, neurointervention, cardiology, and pediatric cardiology were processed using both the AI-based and predecessor image processing algorithms. Thirteen board-certified physicians independently evaluated paired images using five-point Likert scales for overall image quality, contrast of clinically relevant details, noise reduction, and confidence in assessing device and vessel details. The primary endpoint was non-inferiority of overall image quality. Secondary endpoints evaluated superiority across image quality metrics using a predefined hierarchical testing strategy. RESULTS: The AI-based image processing algorithm demonstrated non-inferior overall image quality in 98.9% of all evaluations (95% confidence interval [CI], 98%-100%), meeting the primary endpoint across all imaging modes, anatomic regions, and predefined subgroups. Superiority of the AI-based algorithm was demonstrated for overall image quality, contrast of clinically relevant details, noise reduction, and confidence in assessing device and vessel details across fluoroscopy, roadmap, overlay reference, and native acquisition imaging. Overall image quality superiority was not demonstrated for digital subtraction angiography, likely reflecting the intrinsically higher signal-to-noise characteristics of these acquisitions. Inter-reader agreement analysis demonstrated identical ratings in 49.3% of evaluations, with 94.7% of paired ratings differing by no more than one Likert category. CONCLUSIONS: In this multicenter blinded reader study, AI-based image processing demonstrated non-inferior overall image quality compared with its predecessor across all evaluated imaging modes and patient subgroups, with superior performance for multiple image quality metrics in fluoroscopy, roadmap, overlay reference, and native acquisition imaging. These findings support the clinical utility of AI-based image processing for endovascular imaging and provide a foundation for future studies evaluating its impact on radiation dose and procedural outcomes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Exploring a Novel Processing Algorithm to Improve Image Quality and Reduce X-Ray and Contrast Dose for Coronary Angiography: A Blinded Pilot Study.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Novel Use of an AI-Based Video Interpolation Algorithm to Reduce Radiation Dose in the Cardiac Catheterization Laboratory.Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2026
- Artificial Intelligence for Computer-Aided Detection in Endovascular Interventions: Clinical Applications, Validation, and Translational PerspectivesBioengineering · 2026
- Cardiac-Specific AI Reconstruction Versus Hybrid Iterative Reconstruction for Emergency Coronary CT Angiography: A Paired Single-Center Study Including Triple-Rule-Out CT.Journal of clinical medicine · 2026
- Artificial Intelligence-Derived Fractional Flow Reserve in Routine Clinical Practice: An International Multicenter Retrospective Study.Journal of the Society for Cardiovascular Angiography & Interventions · 2026
- Prospective Shadow-Mode Evaluation of an Artificial Intelligence Tool for Intracranial Aneurysm Detection on CT Angiography: Incremental Yield and Operational Impact.Journal of the American College of Radiology : JACR · 2026