Accuracy of iodinated contrast and saline delivery using a piston-driven and a peristaltic power injector in CT-an in vitro study.
Zahra Alyani Nezhad, Aria M Salyapongse, Courtney J Goetsch, Rachel M Bladorn, Carrie M Bartels, Kelsey L Schluter, Martin G Wagner, Giuseppe V Toia,
Department of Medical Physics, University of Wisconsin-Madison, Madison, WI, USA. · Department of Radiology, University of Wisconsin-Madison, Madison, WI, USA.
Key points
From the paper's conclusion- Both injectors delivered excess ICM while saline flush volumes showed variability. These findings underscore the importance of performance verification to ensure accurate and consistent contrast delivery in clinical practice.
Abstract
OBJECTIVE: To evaluate the accuracy of iodinated contrast media (ICM) and saline flush volumes delivered by a piston- and a peristaltic-based injector in an in vitro study. MATERIALS AND METHODS: Seven clinical injection protocols, ranging from pediatric to bariatric indications and incorporating varying flow rates, ICM volumes, saline volumes, and dilution ratios, were tested. Identical prescriptions were programmed into a piston-based injector and a peristaltic-based injector. Injections were performed into vials. Calibration samples (5-20 mgI/mL) were scanned on a multi-slice CT scanner to generate a calibration curve relating CT number to iodine concentration. The saline test volume and combined ICM + saline flush volume were measured for each injection. CT numbers of mixtures were converted to iodine concentration using the calibration curve, and delivered ICM and saline volumes were calculated. Percent errors between expected and measured volumes were determined. A Wilcoxon signed-rank test assessed statistical significance. RESULTS: For the peristaltic-based injector, ICM volume error ranged from 4.0% to 10.0% and saline flush error from 0.6% to 9.2%. For the piston-based injector, ICM volume error ranged from 9.9% to 40.0% and saline flush error from 1.5% to 50.2%. Differences between expected and delivered volumes were statistically significant across protocols (p = 0.031/= 0.035), except for peristaltic saline flush protocols (Ped 12 mL (p = 0.094), Adult 170 mL (p = 0.156), and High Flow (p = 0.093)). CONCLUSIONS: Both injectors delivered excess ICM while saline flush volumes showed variability. These findings underscore the importance of performance verification to ensure accurate and consistent contrast delivery in clinical practice. KEY POINTS: Question Automated injectors often delivered more contrast than expected. Findings Both systems often underdelivered saline with some instances of overdelivery. Relevance statement Unrecognized variability in contrast and saline delivery across different injection settings may affect enhancement consistency, contrast exposure, and diagnostic reliability. Broader injector performance evaluation across clinically used protocols and clinically simulated setups may improve CT practice.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Accuracy and Precision of Gravity-Driven Infusions.Prehospital emergency care · 2026
- Model-based Pressure-threshold Flow-rate Estimation for Contrast Injection via Upper-arm Central Venous Ports.Academic radiology · 2026
- Performance characterization of syringes used for intravitreal injections: dosing accuracy, particulate contamination, and residual volume.International journal of retina and vitreous · 2026
- Improved liver and portal venous system enhancement with saline flush after contrast injection in portal venous phase computed tomography.Acta radiologica (Stockholm, Sweden : 1987) · 2026
- Evaluation of a personalized total iodine dose (TID) combined with fixed injection duration (FID) protocol using high-concentration (400 mgI/mL) iodinated contrast medium in CT portal venography: a randomized trial.Quantitative imaging in medicine and surgery · 2026
- Repurposing a Dual-Head Programmable Power Injector for Controlled Intra-Arterial Vasodilator Delivery During Cerebral Vasospasm Treatment.World neurosurgery · 2026