Workflow-level implant-position reproducibility in complete-arch scans under scanner-integrated artificial intelligence-based filtering and wireless scanner-to-cradle distance conditions: a device-specific in vitro study.
Maya Iwamoto, Kotaro Saka, Mikio Nakahara, Yoichi Tanaka, Kota Izumida, Junpei Otani, Taketo Miwa, Takahiro Murakami
Clinical Implant Society of Japan, 1-43-9, Komagome, Toshima-ku, Tokyo, 170-0003, Japan. · Kyowa Dental Laboratory, 3-260-1, Shinmatsudo, Chiba, 270-0034, Japan.
Key points
From the paper's conclusion- Within this device-specific workflow, AI-ON was associated with high concordance across the tested conditions. Because only one unit of each scanner and one host computer were evaluated and because wireless quality was not measured directly, the findings do not establish a general AI or distance effect and should not be extrapolated to prosthesis fit or clinical performance.
Abstract
PURPOSE: This device-specific in vitro study compared workflow-level implant-position reproducibility under scanner-integrated artificial intelligence (AI)-based filtering and wired/wireless scanner-to-cradle conditions. METHODS: A maxillary edentulous model with six implants was scanned using one Aoralscan Elite unit, one Aoralscan Elite Wireless unit, and one host computer. AI-based filtering enabled and disabled conditions were defined as AI-ON and AI-OFF, respectively. Five scans were acquired for each of the eight AI × communication-condition combinations: wired reference and Wireless 0.5, 2.5, and 5.0 m (40 datasets). After computer-aided design library matching and virtual-abutment replacement, datasets were registered using three reference bodies. Concordance was the percentage of the virtual-abutment surface area within a |deviation| of ≤ 50 μm of the reference. RESULTS: With AI-OFF, the median concordance was 53.1% at 2.5 m and 50.0% at 5.0 m, compared to 97.7% and 98.7%, respectively, with AI-ON. Exploratory two-way analysis of variance showed an AI × communication-condition interaction [F(3, 32) = 14.92, p < 0.001, partial η² = 0.583], supported by residual-permutation analysis. Holm-corrected exact comparisons showed higher concordance with AI-ON than with AI-OFF in the wired, 2.5-m, and 5.0-m conditions. CONCLUSIONS: Within this device-specific workflow, AI-ON was associated with high concordance across the tested conditions. Because only one unit of each scanner and one host computer were evaluated and because wireless quality was not measured directly, the findings do not establish a general AI or distance effect and should not be extrapolated to prosthesis fit or clinical performance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- In Vitro Comparison of Trueness and Precision of an AI-Driven Real-Time Library Matching Protocol with Irregular Geometry Scan Bodies for Full-Arch Implant ScanningDentistry Journal · 2025
- Effectiveness of an AI-Assisted Digital Workflow for Complete-Arch Implant Impressions: An In Vitro Comparative StudyDentistry Journal · 2025
- Accuracy of artificial intelligence-based functions in intraoral scanners: A systematic review.The Journal of prosthetic dentistry · 2026
- Effect of Non-Splinting Techniques and Non-Calibrated Implant Scan Body Systems on the Accuracy of Complete-Arch Implant Scans: A Comparative Clinical Study.Clinical oral implants research · 2026
- Artificial Intelligence Segmentation Errors in Implant Planning Software Programs: An OverviewClinical Implant Dentistry and Related Research · 2025
- GENERALIZABILITY OF CLOUD-BASED AI SOFTWARE FOR ANTERIOR TOOTH SEGMENTATION IN MULTICENTER CBCT DATASETS: AN EXTERNAL VALIDATION STUDY.Journal of dentistry · 2026