Reproducibility and variability of visual capillary refill time assessment by intensive care physicians: comparison with near-infrared spectroscopy and automated video analysis in a cross-sectional observational study.

Nicole Alejandra Goette, Facundo Javier Gutierrez, Juan Ignacio Mazza, Mario Omar Pozo, Gastón Eduardo Murias

Servicio de Terapia Intensiva, Hospital Británico, Perdriel 74 (C1280AEB), Ciudad Autónoma de Buenos Aires, Argentina. · Servicio de Terapia Intensiva, Hospital Británico, Perdriel 74 (C1280AEB), Ciudad Autónoma de Buenos Aires, Argentina. facundo_gutierrez@yahoo.com.ar.

Observational

Key points

From the paper's conclusion
  • Visual CRT showed moderate reproducibility and did not detect graded perfusion changes. Both NIRS and the Python-based automated software identified condition-related differences that visual assessment missed. These results call for cautious interpretation of isolated CRT when small serial changes inform clinical decisions, and support further evaluation of automated approaches in critically ill patients.

Abstract

BACKGROUND: Capillary refill time (CRT) is a widely used bedside marker of peripheral perfusion in intensive care, and its clinical relevance has grown after recent randomized trials incorporated it into hemodynamic resuscitation algorithms. Reproducibility under routine conditions remains poorly characterized, and objective tools to reduce observer-dependent variability are lacking. We developed a custom Python-based software for automated colorimetric CRT analysis and compared visual assessment with this tool and near-infrared spectroscopy (NIRS) in a controlled experimental model. METHODS: Eight healthy adult volunteers underwent three conditions: baseline, venous congestion at 50% of diastolic blood pressure (DBP), and congestion at 75% DBP. Tissue oxygen saturation (StO₂) was recorded continuously by NIRS every 3 s. Fingertip-pulp compression videos were obtained at 20, 40, 60, 80, and 240 s. Sixteen ICU clinicians assessed 2401 coded videos in random order. The Python colorimetric tool was applied to valid recordings. Reproducibility was evaluated with variance-component models, intraclass correlation coefficients (ICC), coefficients of repeatability (CR), and linear mixed-effects models. RESULTS: Mean visual CRT was 1.77 ± 0.65 s. No significant effects of congestion condition (F = 1.45, p = 0.23) or time (F = 1.48, p = 0.230) were detected; marginal R² was 0.04 and conditional R² was 0.44. Interobserver ICC was 0.52 (95% bootstrap CI 0.42-0.61), with a CR of 1.20 s; mean intraobserver ICC was 0.69 (range 0.42-0.84). Simulations showed ICC rising to 0.76 with three observers, 0.84 with five, and 0.92 with ten. NIRS (904 observations) detected significant condition- and time-related changes (p < 0.001); late-phase StO₂ differed by 8.46% between congestion levels, and nadir StO₂ was lower under greater congestion (p = 0.0018). Automated analysis of 53 recordings yielded a mean CRT of 3.89 ± 1.26 s, a significant congestion effect (F = 6.32, p = 0.016; marginal R² 0.12, conditional R² 0.48), and a mean between-condition difference of 0.68 s (p = 0.017). CONCLUSIONS: Visual CRT showed moderate reproducibility and did not detect graded perfusion changes. Both NIRS and the Python-based automated software identified condition-related differences that visual assessment missed. These results call for cautious interpretation of isolated CRT when small serial changes inform clinical decisions, and support further evaluation of automated approaches in critically ill patients.