Esophageal resonance Raman spectroscopy tracks mixed venous oxygenation in experimental shock and hypoxemia.
Tysen DeWaard, Bonne L Keijer, Franklin Ducatez, Cassidy Ferron, Jason Zhang, Yifeng Peng, Padraic Romfh, John N Kheir
Department of Cardiology, Division of Cardiovascular Critical Care, Boston Children's Hospital, 300 Longwood Avenue, Boston, 02115, MA, USA. · Pendar Medical, LLC, Cambridge, 02138, MA, USA.
Key points
From the paper's conclusion- We evaluated whether a customized probe applied to the esophageal mucosa could provide a continuous measurement of regional capillary hemoglobin oxygen saturation (S Es O 2 ) that tracks changes in mixed venous oxygen saturation for early detection of experimental shock with minimal invasiveness.
Abstract
INTRODUCTION: Mixed venous oxygen saturation (SvO2) reflects the balance between systemic oxygen delivery and consumption and is widely used as an indicator of tissue oxygenation in critically ill patients. However, its measurement requires invasive venous access. We evaluated whether a customized probe applied to the esophageal mucosa could provide a continuous measurement of regional capillary hemoglobin oxygen saturation (SEsO2) that tracks changes in mixed venous oxygen saturation for early detection of experimental shock with minimal invasiveness. METHODS: A miniaturized 405 nm resonance Raman spectroscopy (RRS) probe was developed to quantify esophageal regional capillary hemoglobin saturation (SEsO2) based on spectral decomposition of oxy- and deoxy-hemoglobin Raman peaks. Two rat models were investigated: progressive hemorrhage (n = 12) and graded hypoxia (n = 9). At each stage, SvO2, SEsO2, arterial oxygen saturation (SaO2), and serum lactate were measured. Tracking ability and diagnostic accuracy compared to SvO2 and transabdominal near-infrared spectroscopy (NIRS) were evaluated using linear mixed-effects models (animal random intercepts), (cluster-bootstrapped) receiver operating characteristic curve (ROC) analyses, and Pearson correlation. RESULTS: In both models, baseline SEsO2 was ~20% higher than SvO2, consistent with a regional capillary measurement distinct from mixed venous blood. In hemorrhage, LME showed SEsO2 was positively associated with SvO2 (β = 0.53, 95% CI 0.43-0.62; p < 0.0001) and inversely associated with lactate (β = -0.15, 95% CI -0.18 to -0.12; p < 0.0001). In ROC analysis, SEsO2 discriminated SvO2 < 40% with AUC 0.98 (95% CI 0.95-1.00). For transabdominal NIRS, the AUC was 0.82 (95% CI 0.70-0.95). During graded hypoxia, SEsO2 was also positively associated with SvO2 (β = 0.51, 95% CI 0.43-0.60; p < 0.0001) and inversely associated with lactate (β = -0.11, 95% CI -0.13 to -0.09; p < 0.0001). Discrimination of SEsO2 for SvO2 < 40% was AUC 0.92 (95% CI 0.82-1.00). Transabdominal NIRS demonstrated discrimination with high uncertainty (AUC 0.56, 95% CI 0.23-1.00). CONCLUSIONS: Esophageal RRS enables molecularly specific measurement of regional capillary hemoglobin oxygen saturation and thereby tracks SvO2 across distinct experimental shock paradigms. These preclinical findings support further investigation of esophageal RRS as a potential minimally invasive monitoring approach for detection of systemic oxygen imbalance.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Continuous Site-Specific Intravascular Monitoring of Arterial and Venous Blood Oxygenation In Vivo.Physiological measurement · 2026
- Controlled systemic hypoxia induces consistent and reversible changes in retinal tissue oxygen saturation measured by ocular oximetry.Microvascular research · 2026
- Real-Time SpO2 Estimation from Chest Reflectance Photoplethysmography: Algorithm Development and Clinical Validation Against Arterial SaO2.Sensors (Basel, Switzerland) · 2026
- Feasibility of the ForeSight tissue oximetry system for estimating cerebral and renal tissue oxygen saturation in anesthetized dogs during hyperoxia, normoxia, and hypoxia.American journal of veterinary research · 2026
- Development and validation of a reflectance-based approach for oxygen saturation estimation using high-speed hyperspectral imaging.Proceedings of SPIE--the International Society for Optical Engineering · 2026
- Association of oxygen reserve index with different nasal oxygen flow rates during sedated gastrointestinal endoscopy: A retrospective cohort study.Therapeutic advances in gastroenterology · 2026