Development and validation of a GC-MS method for the determination of ethylene glycol and diethylene glycol in pharmaceutical syrups.
Muhammad Amirul Amil, Muhammad Shahariz Mohamad Adzib, Nur Baizura Bujang, Kok Zheng Gan, Sara Khalida Jamal Khair, Malissa Wani Mohammad Amini, Ahmad Zariq Syafiq Zaire, Ahmad Yusri Mohd Yusop
Centre of Compliance and Quality Control, National Pharmaceutical Regulatory Agency, Ministry of Health Malaysia, Petaling Jaya, Selangor, 46200, Malaysia. amirulamil@npra.gov.my. · Centre of Compliance and Quality Control, National Pharmaceutical Regulatory Agency, Ministry of Health Malaysia, Petaling Jaya, Selangor, 46200, Malaysia.
Key points
From the paper's conclusion- Therefore, this study emphasises the importance of regulatory surveillance, supported by validated analytical testing, to ensure the quality and safety of pharmaceutical syrups.
Abstract
Contamination of pharmaceutical syrups with ethylene glycol (EG) and diethylene glycol (DEG) has led to serious poisoning incidents worldwide, highlighting the necessity for strong regulations and reliable testing methods. A gas chromatography-mass spectrometry (GC-MS) method was developed and validated in accordance with ICH Q2(R2) guidelines. The method was developed for pharmaceutical syrup matrices and validated using multiple commercially available products, supporting its applicability for routine regulatory surveillance. It exhibited excellent linearity, with r² of 0.9991 for EG and 0.9963 for DEG. The detection (LOD) and quantification (LOQ) limits were 0.02% v/v and 0.06% v/v, respectively, below the regulatory limit of 0.10% set by the National Pharmaceutical Regulatory Agency (NPRA), Ministry of Health Malaysia. Accuracy ranged from 94.2% to 108.4%, and the %RSD for intermediate precision was 1.4% for EG and 2.5% for DEG, confirming the method's reproducibility. When the validated method was applied to 51 pharmaceutical syrup samples available in the Malaysian market as part of post-market surveillance, 13 samples were detected to have EG. Of these, 11 samples had EG concentrations between the LOD and the LOQ, while two samples (PS-13 and PS-26) had measurable levels at 0.09% v/v. DEG was not detected in any sample. Therefore, this study emphasises the importance of regulatory surveillance, supported by validated analytical testing, to ensure the quality and safety of pharmaceutical syrups. Strengthening local surveillance mechanisms will further strengthen NPRA's regulatory capacity and safeguard public health.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Rapid screening of ethylene glycol and diethylene glycol in raw materials and medicinal syrups using low-cost field deployable assaysScientific Reports · 2025
- Optimization of a SWIEET-LC-MS/MS method for quantification of regulated pyrrolizidine alkaloids: Application to honey samples.Talanta · 2026
- Development and validation of the analytical method for determination of residual solvents (2-propanol and n-hexane) in the drug substance Inosine Pranobex.BMC chemistry · 2026
- Evaluation and Measurement of Potential Genotoxic Impurities in Empagliflozin Active Pharmaceutical Ingredient Using Tandem Liquid Chromatography Mass Spectrometer, Accompanied by an Advanced Greenness AssessmentSeparation Science Plus · 2026
- Targeted HPLC-MS/MS Profiling of Diacylglycerols for Krill Oil Authentication and Adulteration DetectionbioRxiv · 2026
- Development and Validation of a GC-MS/MS Method for NDMA Determination in Furazidine-Containing Drug Products: Analytical Evaluation of Precursors and Risk Assessment.International journal of molecular sciences · 2026