Green chiral HPLC method on a polysaccharide column under hydro-organic conditions for the determination of Levobupivacaine, Ropivacaine, and their enantiomeric impurities.
Nehal G Sedki, Ahmed S Fayed, Nahla N Salama, Maha A Hegazy
Postgraduate program in Pharmaceutical Analytical Chemistry, Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Cairo University, Kasr El-Aini Street, Cairo, ET-11562, Egypt. nehalsedki82@gmail.com. · Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Cairo University, Kasr El-Aini Street, Cairo, ET-11562, Egypt.
Key points
From the paper's conclusion- To our knowledge, the developed method is the first application of hydro-organic conditions on a polysaccharide column for these local anesthetics, offering a rapid, selective, and sustainable alternative to traditional modes used in chiral LC methods.
Abstract
A novel, rapid, green, and enantioselective HPLC method was developed and validated using a polysaccharide-based chiral column for the individual determination of Levobupivacaine and S (-) Ropivacaine together with their respective chiral impurities (R (+) Bupivacaine and R (+) Ropivacaine). The separation was optimized on a Chiralpak AD column under hydro-organic conditions using an isocratic mobile phase of isopropanol and water (85:15, v/v) at flow rate 0.5 mL/min, with UV detection at 220 nm. The method was validated according to ICH guidelines and assessed for environmental impact using NEMI, Analytical Eco-Scale, GAPI, and AGREE tools. Baseline separation of each enantiomer pair was achieved within 8 min. Good linearity (R2 > 0.999) was exhibited over 20-200 μg/mL for the S-enantiomers and 2 - 20 μg/mL for the R-enantiomers. Precision (RSD < 1.36%) and accuracy (recoveries 99.4-100.7%) were satisfactory. Detection and quantitation limits for the chiral impurities were < 0.7 μg/mL and < 2.1 μg/mL, respectively. Thermodynamic studies indicated an enthalpy-driven enantioseparation involving hydrogen bonding, π-π interactions, and steric hindrance. The greenness assessment confirmed a superior environmental profile compared to literature methods, with an Eco-Scale score of 92 and AGREE score of 0.70. To our knowledge, the developed method is the first application of hydro-organic conditions on a polysaccharide column for these local anesthetics, offering a rapid, selective, and sustainable alternative to traditional modes used in chiral LC methods. Its simplicity and robustness make it well suited for routine quality control applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Green Analytical Chemistry-Oriented RPLC Method for Determining the Dissociation Constant of the Nootropic Drug Piracetam in Binary Solvent SystemsJournal of Chemical & Engineering Data · 2026
- Development and Validation of Chiral Supercritical Fluid Chromatographic Method for Rapid Separation of Viloxazine Using an Immobilized Amylose‐Based Chiral Stationary PhaseChirality · 2026
- Ecologically conscious profiles assessments for economic, and concurrent HPLC detection of chlordiazepoxide and clidinium bromide in bulk and tablets solid dosage formResults in Chemistry · 2025
- Validated stability indicating reverse phase liquid chromatography method for oxcarbazepine and its degradation products with greenness evaluation.Scientific reports · 2026
- Stability indicating HPTLC method for simultaneous analysis of sitagliptin phosphate monohydrate and pioglitazone hydrochloride with greenness assessment using AGREE, ESA, BAGI and CACIDiscover Chemistry. · 2026
- Stability indicating green HPLC method for Imeglimin hydrochloride determination in pharmaceutical tablets with comprehensive assessmentScientific Reports · 2026