Personalized FES-cycling patterns via optimal control: from musculoskeletal simulation to experimental validation.
Tiago Coelho-Magalhães, Christine Azevedo-Coste, Henrique Resende-Martins, François Bailly
CAMIN, Inria, Univ Montpellier, 860 Rue Saint Priest, Montpellier, 34095, France. · Graduate Program in Electrical Engineering, Universidade Federal de Minas Gerais, Av. Antônio Carlos 6627, Belo Horizonte, 31270-901, Minas Gerais, Brazil.
Key points
From the paper's conclusion- We demonstrate the feasibility of obtaining a stimulation pattern via trajectory optimization to track a 3.6 rad/s (35 rpm) cadence and a 25 W crank power output by optimizing muscle stimulation pulse durations.
Abstract
This study introduces an optimal control framework for functional electrical stimulation (FES) cycling that produces personalized stimulation patterns directly transferable from simulation to experimental cycling, without the need for empirical tuning. We use a physiological muscle model that accounts for the effects of electrical stimulation on muscle contraction dynamics, whose parameters are identified using experimental data obtained from a dedicated protocol conducted with a participant with spinal cord injury (SCI). We demonstrate the feasibility of obtaining a stimulation pattern via trajectory optimization to track a 3.6 rad/s (35 rpm) cadence and a 25 W crank power output by optimizing muscle stimulation pulse durations. The simulations are conducted using two different models-one comprising six muscles and the other four-to explore different experimental conditions. Finally, we assess whether the resulting stimulation patterns enable the participant to pedal a recumbent tricycle without further modification. The generated patterns successfully produce continuous, coordinated pedaling. The average power output was 18 W with a mean angular velocity of 3.93 ± 0.80 rad/s for the six-muscle configuration, and 19 W with a mean angular velocity of 3.90 ± 1.13 rad/s for the four-muscle configuration. This study introduces a physiological model-based control approach for FES cycling, bridging musculoskeletal simulation and experimental validation, and underscores the potential of predictive simulations and optimal control to personalize FES-assisted protocols.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- From simulation to stimulation: a predictive synergy-based control strategy for functional electrical stimulation cycling in a person with spinal cord injury.Journal of neuroengineering and rehabilitation · 2026
- A kinematic-and-muscular modulation strategy for FES-assisted upper limb rehabilitation: a feasibility studyScientific Reports · 2025
- Feasibility and apparent safety of functional electrical stimulation-assisted cycling in amyotrophic lateral sclerosis and primary lateral sclerosis: a single-arm longitudinal pilot feasibility study.European journal of translational myology · 2026
- Functional electrical stimulation-assisted cycling therapy with virtual reality biofeedback for functional recovery in people with chronic spinal cord injury: a feasibility case series.Spinal cord series and cases · 2026
- Closed-loop neuroprosthetic control through spared neural activity enables proportional foot movements after spinal cord injury.Journal of neural engineering · 2026
- Adaptive weighted fuzzy ILC for personalized FES trajectory reconstruction using an anatomical prior.Journal of neural engineering · 2026