Characterisation and management of therapy-interrupting system errors associated with the use of the HomeChoice cycler for automated peritoneal dialysis.
Annemarie Albert, Stefan Richter, Rainer Peter Woitas, Philipp Stieger, Rüdiger C Braun-Dullaeus, Christian Albert
Diaverum Renal Services, Am Neuen Garten 11, 14469, Potsdam, Germany. · Diaverum Deutschland GmbH, Hansastraße 23, 80686, München, Deutschland.
Key points
From the paper's conclusion- These findings may be useful to guide training purposes for nephrology personnel and persons doing APD. For the manufacturer, the information provided may be helpful to improve error management and information transmission to the persons doing APD with the HomeChoice devices.
Abstract
BACKGROUND: Methodological and technical education of nurses is required to address challenges of technology integration in peritoneal dialysis (PD). We analysed the nature of therapy interrupting system errors in automated peritoneal dialysis (APD) cyclers of a German PD reference care centre and provide applicable resolution proposals. METHODS: We documented therapy interrupting system errors of HomeChoice Pro and HomeChoice Claria cyclers (Baxter, BX, Unterschleißheim, Germany) including the years 2015-2023. We systematically characterise and suggest resolution strategies to these errors based on routinely collected health data and user experience to safely guide nephrology professionals and patients during troubleshooting. The 'reporting of studies conducted using observational routinely collected health data' (RECORD)-statement was considered. RESULTS: We identified 49 different therapy interrupting system errors which were assigned to eight major categories: Errors associated to the pressure chamber (N=18), the Digitalboard/EEPROM (N=18, Electrically Erasable Programmable Read Only Memory), air in set (N=3), opening of the set's door (N=3), those associated to the energy system (N=3), the heater system (N=3) and a variety of errors resulting in an abrupt termination or non-commencement of the treatment session without clear group assignment (undefined, N=1) or without sufficient documentation. In total, 138 error incidents were documented during the observational timeframe. In frequential order errors were attributed to the pressure chamber (21.74%,), and air in set (19.57%) where specifically error 2240 occurred in 17.39% of total cases, the Digitalboard/EEPROM (17.39%), the energy or battery (16.67%; No. 1032 in 19/23 cases), the set's door (7.25%) and the heater (4.35%). In 17 cases (12.32%) no error number was documented by the patient or staff. One error cause was registered as undefined (0.72%). Overall, errors were rarely occurring in <0.25% of treatment sessions performed in the observational timeframe. CONCLUSION: These findings may be useful to guide training purposes for nephrology personnel and persons doing APD. For the manufacturer, the information provided may be helpful to improve error management and information transmission to the persons doing APD with the HomeChoice devices.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Patient-Centered Observations of Outpatient Hemodialysis: An Exploratory Systems Approach to Understanding Infection Prevention Using Human Factors.Kidney360 · 2026
- Human factors testing in a novel automated peritoneal dialysis system.Peritoneal dialysis international : journal of the International Society for Peritoneal Dialysis · 2026
- From Remote Monitoring to Actionable Care: Real-World Implementation of an Interoperable Digital Ecosystem for Peritoneal Dialysis.Healthcare (Basel, Switzerland) · 2026
- Development and preliminary evaluation of a mobile-based situational simulation retraining system for patients undergoing peritoneal dialysis: A pilot study.Digital health · 2026
- Standardizing the digital frontier: A Nordic survey of remote patient management practice patterns and variations in peritoneal dialysis care.Peritoneal dialysis international : journal of the International Society for Peritoneal Dialysis · 2026
- Perspective: The Future of Peritoneal Dialysis in the Era of Telemedicine and Artificial Intelligence.Blood purification · 2026