Retention behavior and surface alterations of ball and TRI raptor unsplinted overdenture attachments after thermomechanical aging: an in vitro study.
Egemen Sezen, Huseyin Kurtulmus
Department of Prosthetic Dentistry, Faculty of Dentistry, Istanbul Yeni Yuzyil University, Sutluce Mahallesi, Binektası Sokak. No:10, Beyoglu, Istanbul, 34445, Turkey. dregemensezen@gmail.com. · Department of Prosthetic Dentistry, Faculty of Dentistry, Izmir Tinaztepe University, Izmir, Turkey.
Key points
From the paper's conclusion- Retentive-force loss was not uniform across the tested unsplinted overdenture attachment subgroups. Insert-specific material and design characteristics influenced retention behavior under in vitro aging conditions. Subgroup rankings within the TRI Raptor Smart Box system should be interpreted with caution, as an order effect arising from sequential patrix use cannot be excluded.
Abstract
BACKGROUND: Retention loss is a recognized maintenance concern in implant overdenture therapy and is closely associated with progressive matrix wear in unsplinted attachment systems. Despite growing evidence on conventional ball and Locator attachments, comparative data on the TRI Raptor Smart Box system under combined thermomechanical aging conditions remain limited. This study aimed to compare the retentive-force behavior and wear-related surface alterations of TRI ball and TRI Raptor Smart Box unsplinted overdenture attachment systems after artificial aging. METHODS: In this in vitro study, a ball attachment system (clear and gray matrix inserts) and a TRI Raptor Smart Box system (purple, clear and pink inserts) were evaluated (n = 10 per subgroup). Custom specimen holders were fabricated to simulate attachment-matrix engagement. Artificial aging was performed in a thermocycling chewing simulator between 5 °C and 55 °C with simultaneous cyclic loading. Retentive-force measurements were obtained after 120, 720, 1440 and 2160 cycles, corresponding approximately to 1, 6, 12 and 18 months of clinical use. Pull-off force was measured using a microtensile testing device. Representative components were examined by scanning electron microscopy after the final aging interval. RESULTS: Retentive force decreased significantly over time in all subgroups. In the ball attachment system, the gray insert consistently showed higher retentive-force values than the clear insert (12.55 ± 0.21 N vs. 11.58 ± 0.37 N at 120 cycles; 2.93 ± 0.59 N vs. 2.25 ± 0.42 N at 2160 cycles). In the TRI Raptor Smart Box system, the purple insert showed the highest retentive-force values at 1, 6 and 18 months, with a transient crossover at 12 months, when the clear insert showed the highest mean value (10.36 ± 0.38 N vs. 9.06 ± 0.53 N), however, an order effect cannot be excluded given the sequential patrix use. The pink Raptor insert showed the lowest values overall, reaching 1.67 ± 0.31 N at 2160 cycles. SEM demonstrated subgroup-specific wear-related surface alterations after cyclic aging. CONCLUSIONS: Retentive-force loss was not uniform across the tested unsplinted overdenture attachment subgroups. Insert-specific material and design characteristics influenced retention behavior under in vitro aging conditions. Subgroup rankings within the TRI Raptor Smart Box system should be interpreted with caution, as an order effect arising from sequential patrix use cannot be excluded.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Effect of Mechanical and Thermal Aging on the Retention Stability of Attachment Systems for Mandibular Implant Overdentures.Journal of functional biomaterials · 2026
- Retentive Force Assessment and Wear Exploration of CAD-CAM Fabricated PEEK, Zirconia, and PMMA Mini Ball Attachments for Tooth-Supported Overdenture: A Laboratory Study.Dentistry journal · 2026
- Retention and patient-reported outcomes of low-profile attachment system for mandibular overdentures: A crossover clinical trial comparing 3D-printed and conventional dentures.Journal of prosthodontics : official journal of the American College of Prosthodontists · 2026
- Research progress on the effects of implant-supported overdentures on residual ridge remodelling in edentulous jaws: a narrative review.Acta odontologica Scandinavica · 2026
- Round versus Locator Attachment Effect on The Retention Forces of Implant-Supported Maxillary Overdentures. A Comparative In Vitro StudyRazi Medical Journal · 2026
- Clinical performance and prosthetic outcomes in reinforced implant-supported overdentures: A systematic review.The Journal of prosthetic dentistry · 2026