Timing of mechanical CPR deployment and defibrillation process quality in shockable cardiac arrest: a multicenter retrospective cohort study.
Shao-Jie Li, Wei Wang, Xiao Li, Jia Qiao
Emergency Department, Qingdao Municipal Hospital Group, 1 Jiaozhou Road, Shibei District, Qingdao, 266000, China. · Emergency Department, Qingdao Municipal Hospital Group, 1 Jiaozhou Road, Shibei District, Qingdao, 266000, China. lixiao0428@sohu.com.
Key points
From the paper's conclusion- Pre-shock mCPR activation was associated with substantial delay and interruption in the initial defibrillation sequence. In patients with a shockable rhythm, mCPR should not be initiated before the first shock when effective manual compressions can be maintained; initial defibrillation should be completed first.
Abstract
BACKGROUND: Mechanical cardiopulmonary resuscitation (mCPR) may support consistent compressions but could interfere with early defibrillation when deployed before the first shock. We examined the association between pre-shock mCPR activation and defibrillation-process quality in adults with an initial shockable rhythm. METHODS: This multicenter retrospective cohort study included adults resuscitated between January 2022 and December 2024. Exposure was defined as manual CPR until the first shock versus mCPR active before the first shock. The primary outcome was CPR-to-first-shock time; secondary outcomes captured rhythm-analysis-to-shock timing, shock-related pauses, chest compression fraction, and prespecified delay thresholds. Multivariable regression with marginal standardization was used, with propensity-score weighting, complete-cohort, alternative-threshold, adjudication, operational-complexity, and OHCA/IHCA interaction analyses assessing robustness. RESULTS: Among 486 patients, 386 (79.4%) received manual CPR until the first shock and 100 (20.6%) had pre-shock mCPR. Median CPR-to-first-shock time was 112.2 seconds (Q1-Q3, 91.9-132.1) and 201.6 seconds (178.8-220.7), respectively. Pre-shock mCPR was associated with a longer CPR-to-first-shock interval (adjusted time ratio, 1.83; 95% CI, 1.74-1.93) and an adjusted absolute difference of 90.8 seconds (95% CI, 82.1-100.8). Rhythm-analysis-to-shock time, pre-shock pause, and peri-shock pause were also longer, whereas chest compression fraction was lower. The adjusted probability of first-shock delay >180 seconds was 3.0% with manual CPR and 75.9% with pre-shock mCPR. There was no strong evidence of OHCA/IHCA effect modification for the primary outcome (P interaction=0.080). Sensitivity analyses were consistent with the primary findings. CONCLUSIONS: Pre-shock mCPR activation was associated with substantial delay and interruption in the initial defibrillation sequence. In patients with a shockable rhythm, mCPR should not be initiated before the first shock when effective manual compressions can be maintained; initial defibrillation should be completed first. CLINICAL TRIAL NUMBER: Not applicable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A randomised controlled trial of defibrillation with manual pressure augmentation during out-of-hospital cardiac arrestResuscitation · 2026
- The association of mechanical versus manual chest compressions with neurological outcomes at hospital discharge among out-of-hospital cardiac arrests treated with intra-arrest transportResuscitation · 2026
- Mechanical CPR - Assisted Percutaneous Coronary Intervention During Cardiac Arrest: Feasibility and Outcomes.Medical principles and practice : international journal of the Kuwait University, Health Science Centre · 2026
- Role of mechanical circulatory support in mechanical complications of acute myocardial infarction.REC, interventional cardiology · 2026
- Preemptive mechanical circulatory support in ventricular tachycardia ablation: A systematic review and meta-analysis.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Impella versus intra-aortic balloon pump in resuscitated out-of-hospital cardiac arrest patients with coronary artery diseaseEuropean Heart Journal · 2026