The Failure Index as Marker of the Neural Health Profile in MED-EL CI Users: A Retrospective Analysis.
Wiebke Konerding, Cornelia Batsoulis, Peter Baumhoff, Heval Benav, Lutz Gärtner, Annette Günther, Onhintz de Olano Dieterich, Daniel Schurzig,
Department of Experimental Otology, Hannover Medical School, Hannover, Germany. · MED-EL Research Center, MED-EL Medical Electronics GmbH, Hannover, Germany.
Key points
From the paper's conclusion- These results were in line with the hypothesis that the FI may serve as a clinical tool to identify implanted ears with reduced neural health and to identify contacts stimulating areas of reduced SGN survival and integrity.
Abstract
Cochlear implants (CIs) enable hearing via direct electrical stimulation of the spiral ganglion neurons (SGN). Outcomes with a CI depend, in part, on the number and excitability of the SGNs. In an animal model, we introduced the Failure Index (FI) as an electrically-evoked compound action potential (eCAP)-derived marker of cochlear health. The FI informs about the presence, site, and size of SGN lesions. Here, we translated the FI to clinical recordings from human MED-EL CI users. In this retrospective study, we included patient data recorded between 2016 and 2024 from 199 ears with postlingual hearing loss and 79 ears with prelingual hearing loss. The averaged FI values over all contacts of the array were stable within the analysis period (3rd month to 1st year postoperatively). The FI increased with age and was elevated in etiologies associated with higher SGN loss. Using 3D reconstruction from cone-beam computed tomography scans, we confirmed that the FI was independent of electrode-to-modiolus distance (0.1-2.5 mm). The FI showed individual patterns along the array, with maxima usually found at basal contacts, corresponding to reduced neural health at high-frequency cochlear regions. In a selected group of postlingual-deaf ears, we found higher correlation coefficients between speech-perception scores and the FI compared to other eCAP-derived markers (i.e., threshold, slope, and amplitude). These results were in line with the hypothesis that the FI may serve as a clinical tool to identify implanted ears with reduced neural health and to identify contacts stimulating areas of reduced SGN survival and integrity.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- The Interphase Gap Effect in Cochlear Implant Users: Biological Basis, Parameter Selection, Analytical Methods, and Quantitative ScalesJournal of the Association for Research in Otolaryngology · 2026
- A rapid, machine-learning readout of cochlear neural degeneration that transfers from animals to humans.bioRxiv : the preprint server for biology · 2026
- Establishing reliable neural health measures in adult cochlear implant users.Frontiers in neuroscience · 2026
- Electrically evoked compound action potentials in adult and pediatric cochlear implant users: I. Cochlear nerve responsiveness and electrode-neuron interface index.Hearing research · 2026
- Novel psychoacoustic measure of spectral-temporal processing for discrete regions of the cochlear array: a proof-of-concept study.Audiology & neuro-otology · 2026
- The Effect of Etiology, Array Type and Duration of Deafness on Focused Detection Thresholds in Cochlear Implant Listeners.Trends in hearing · 2026