Sintered titanium-bioceramic composites: A finite element performance analysis suggested for use in cranial surgical fixation plates.
Anthony O Ogunmefun, Festus O Fameso, Moipone L Teffo, Lucey M Mavhungu, Emmanuel R Sadiku, Olusegun M Ilori, Akintunde Babatunde, Snyman Jacques
Department of Chemical, Metallurgical, and Materials Engineering, Tshwane University of Technology, Pretoria, South Africa. · Center for Climate Change, Water Security and Disaster Management, Department of Civil Engineering, Tshwane University of Technology, Pretoria, South Africa.
Key points
From the paper's conclusion- This computational model design and the result are notable for the mechanical performance validation of sintered titanium-bioceramic composites for biomedical applications.
Abstract
Optimizing titanium-bioceramic composites via the SPS technique, and using FE analysis to assess their mechanical performance, forms the basis of this study, which aims at replacing the conventional Ti-alloy frequently used in craniotomy procedures. Composites TB1, TB2, and TB3 are formed with uniform volume percentage Ti-alloy and varied percentages of Zirconia (1.5,1.0, 0.5) vol.%, and constant 0.5 vol.% Si3N4 for all samples. With 1000°C temp, 50 MPa pressure,15 minutes dwell time, and 100°C/min heating rate, the powders were consolidated. The microstructural evolution and bulk morphology of the starting powders and sintered discs produced were investigated using SEM-EDS. The hardness and elastic modulus of the trio-composites were also investigated using the nanoindentation technique. All composite samples attained theoretical densities of 97.24, 98.27, and 99.37 percent, respectively, well above 96.62% of the unreinforced sintered Ti-alloy. Composite TB3 recorded the highest in relative density, elastic modulus, Vickers' hardness, and Nano hardness at 99.37%, 172.75 GPa, 795.77 MPa, and 8941.75 MPa, respectively. TB3 was chosen for the mechanical validation process through a computerized 3-dimensional FEA model, utilizing the front-end commercial software via the Complete Abaqus Environment (CAE) with robust finite element analysis software. The radial mechanical and coupled radial thermomechanical loadings concept, crucial for understanding the mechanical performance of the sintered plate, reveals elastic deformation rather than plastic deformation, suggesting that 'yielding' will not occur under the applied loads. This computational model design and the result are notable for the mechanical performance validation of sintered titanium-bioceramic composites for biomedical applications.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- New developments of biomedical porous titanium alloys prepared by spark plasma sinteringMetals Advances · 2026
- Design and comparison of custom cranial prostheses made of titanium and polyether ether ketone using the finite element methodJournal of Materials Science: Materials in Medicine · 2026
- Benchmarking PA12 and PA12CF35 for selective laser sintering of patient-specific implants: a thermo-mechanical analysisJournal of the mechanical behavior of biomedical materials/Journal of mechanical behavior of biomedical materials · 2025
- Biomechanical Performance of Zirconia–Calcium Silicate–Silver Hybrid Dental Crown Under Static and Transient Dynamic Loading: A Finite Element and TOPSIS ‐Based Multi‐Criteria EvaluationJournal of Biomedical Materials Research Part B Applied Biomaterials · 2026
- Next-generation craniomaxillofacial implants for reconstructive surgery: balancing biomechanics, biocompatibility, and bioactivityInternational Journal of Oral Science · 2026
- Microstructural Evolution, Multi‐Scale Strengthening and Mechanical Properties of TiC‐Based Composites via Reactive SinteringJournal of the American Ceramic Society · 2026