Influence of Infusion Duration and Tubing Length on Insulin Lispro Stability in Subcutaneous Insulin Pump Delivery Systems.
Isabella F de Luna, Ali Bozorgnezhad, Jackie Cullinan, Martha Essandoh, Diaa Shakleya, Maxwell Korang-Yeboah, Ashwinkumar Bhirde, Suvajyoti Guha
Office of Pharmaceutical Quality Research, Office of Pharmaceutical Quality, Center for Drug Evaluation and Research, U.S. Food and Drug Administration, Silver Spring, MD, USA. · Division of Applied Mechanics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, Silver Spring, MD, USA.
Key points
From the paper's conclusion- Flow imaging proved more sensitive than light obscuration for this in vitro analysis. Polyethylene shed from infusion tubing was identified as a source of particulate matter. These findings suggest that sensitive analytical methods may offer utility in characterizing infusion set systems.
Abstract
BACKGROUND: Subvisible particulates in insulin formulations pose risks because of their potential to cause pump occlusion and immune responses in patients during subcutaneous delivery through insulin pump systems. The impact of the length of the infusion set tube and the duration of infusion on particulates has not been fully characterized. Using insulin lispro as a model product, this in vitro study evaluated physical and chemical characterization methods for detecting subvisible particulates. METHOD: Insulin lispro was infused continuously at 0.2 U/h for 3 and 8 days using 2 pump-infusion set systems with tubing lengths ranging from 18 to 43 inches. Particulate characterization using flow imaging, chemical identification using morphologically directed Raman spectroscopy, and monomer and excipient characterization using size exclusion chromatography and gas chromatography-mass spectrometry were performed. RESULTS: Flow imaging microscopy revealed that particulate concentrations were higher in infused samples compared with uninfused controls, with the greatest increase observed with the longest tubing lengths and infusion durations. This method detected a significantly higher concentration of particulates (P < .05) than the compendial light obscuration method. Morphologically directed Raman spectroscopy identified silicone oil and polyethylene as primary constituents of particulates above 10 µm, indicating leaching of silicone oil from syringes and polyethylene shedding from infusion set tubing walls. CONCLUSIONS: Flow imaging proved more sensitive than light obscuration for this in vitro analysis. Polyethylene shed from infusion tubing was identified as a source of particulate matter. These findings suggest that sensitive analytical methods may offer utility in characterizing infusion set systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a method for vancomycin monitoring in interstitial fluid using a novel porous microneedle array patch.Drug metabolism and pharmacokinetics · 2026
- Protein aggregation in self-lubricating piston pumps: The impact of sustained shear stress.International journal of pharmaceutics · 2026
- Peripheral intravenous waveform analysis - a novel tool to confirm intravascular catheter placement.Journal of clinical monitoring and computing · 2026
- In vitro Evaluation of the Physicochemical Stability of Multi-Drug Y-Site Admixtures Using Dynamic Monitoring and Chemometrics.Drug design, development and therapy · 2026
- Accelerated dissolution testing conditions of long-acting injectable suspensions: an exploration of dissolution rate and particle behaviour through in vitro imaging and in silico methods.European journal of pharmaceutical sciences : official journal of the European Federation for Pharmaceutical Sciences · 2026
- Stability Testing of Intraventricular Irrigations Diluted in Lactated Ringer's Solution.International journal of pharmaceutical compounding · 2026