Closing the gap in ultrasound probe hygiene: Improving ultrasound probe decontamination through evidence-based interventions.
Jincy Jerry, Michelle Cranmer, Vasantha Gopalan, Trevor Duffy, Mary Walsh, Sine Gilchriest, Paul Ryan, Ana Safara,
NHS Ayrshire & Arran, Kilmarnock, UK. · Mater Misericordiae University Hospital, Dublin, Ireland.
Key points
From the paper's conclusion- In all six areas that were fully subject to the interventions and re-audit, there was improvement.
Abstract
BACKGROUND AND AIM: Ultrasound is integral to many aspects of healthcare. Patient harm can occur when ultrasound decontamination is not adequate. Evidence-based guidance exists to inform operators of appropriate practice. A clinical audit was performed to assess compliance with these standards in an acute hospital. METHODS: An assessment tool was developed to capture data across multiple domains to evaluate compliance with ultrasound reprocessing standards. Key variables included operator identity, professional role, and location, alongside procedure type and Spaulding classification to ensure correct risk stratification. Infection prevention and control nurses observed ultrasound procedures, categorising probes as critical, semi-critical or non-critical and assessed whether the correct level of decontamination was applied. Adherence to manufacturer instructions for use and local standard operating procedures was evaluated. Traceability of reprocessing was examined. Compliance was summarised as proportions by year and by Spaulding category with statistical analysis performed in Python. Between audit cycles, a multi-modal quality improvement plan was devised and implemented. RESULTS/DISCUSSION: Between the pre- and post-intervention phases, adherence to appropriate disinfection selection and technique improved significantly overall (p < 0.0001) and specifically in critical and non-critical probes. In all six areas that were fully subject to the interventions and re-audit, there was improvement. In the one area subject to re-audit where staff did not participate in training programmes, there was a decline in compliance, supporting that the availability of a standard operating procedure and appropriate equipment is not sufficient on its own but that training of operators is a key contributing factor to improvement.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Ultrasound probe disinfection practice and associated factors among radiologists and radiology residents in Ethiopia, 2025: a cross-sectional study.Antimicrobial resistance and infection control · 2026
- Evidence-based preprocessing improvement bundle for surgical instruments: reducing surgical infection risks via quality improvement.Frontiers in public health · 2026
- Inactivation of microorganisms in the complex regions of transvaginal ultrasound probes by a UVC-LED light based disinfection system.PloS one · 2026
- UV-C disinfection of endocavitary ultrasound probes against Candidozyma auris : a rapid and chemical-free solutionAntimicrobial Stewardship & Healthcare Epidemiology · 2026
- Microbial contamination and decontamination of reusable digital intraoral imaging equipment: a narrative review.Frontiers in oral health · 2026
- Surface-associated Candida auris recovery during five intensive care unit cluster responses: a retrospective descriptive study with structured evidence mapping.Frontiers in cellular and infection microbiology · 2026