Dual-Plate Allograft Reconstruction for Expansile Lytic Lesions of the Metacarpal May Allow for Immediate Postoperative Weightbearing in a Biomechanical Setting.
Mathangi Sridharan, Alyssa Tomkinson, Mikayla Mefford, Kasra Rahmati, Tyler Clites, Lauren E Wessel
Department of Orthopaedic Surgery, University of California Los Angeles, USA. · Departments of Mechanical and Aerospace Engineering and Bioengineering, University of California Los Angeles, USA.
Key points
From the paper's conclusion- Dual-plate fixation constructs for allograft reconstruction of the metacarpal are comparable in mechanical stability to standard metacarpal fracture fixation that allow for immediate postoperative range of motion.
Abstract
BACKGROUND: This biomechanical study investigates whether reconstruction of expansile lesions of the metacarpal with fibular allograft and a dual-plate construct is strong enough for immediate postoperative weightbearing. METHODS: Twelve cadaveric middle finger metacarpals (6 matched pairs) from adult donors were used. One side of each pair was randomly assigned to undergo allograft reconstruction with dual plating (experimental construct). The matched contralateral side was assigned to the control construct, a transverse fracture with standard, 8-hole dorsal plating. In the experimental construct, an expansile bone lesion was simulated in the native metacarpal by resecting 1.5 cm from the proximal carpometacarpal joint to 1.5 cm from the distal articular surface and was reconstructed with fibular allograft. Biomechanical testing was performed by a 6-degree-of-freedom robotic KUKA manipulator. The distal portion of each specimen was clamped in place, and a cantilever bending force was applied by the KUKA to test the load to failure. Failure was defined as periprosthetic fracture, fracture displacement, or screw pullout. Paired t tests were used to compare load to failure between matched pairs (R studio). Significance was set at P < .05. RESULTS: The load to failure of dual-plate allograft constructs did not significantly differ from that of the control construct, 494.7 ± 239.07 N (95% CI, 243.81-745.59 N) versus 342.2 ± 99.9 N (95% CI, 237.45-447.04 N), P = .19. Experimental constructs failed at the proximal allograft-native bone interface. Control constructs failed via fracture displacement. CONCLUSIONS: Dual-plate fixation constructs for allograft reconstruction of the metacarpal are comparable in mechanical stability to standard metacarpal fracture fixation that allow for immediate postoperative range of motion.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Intercalary Allograft Reconstruction of the Femur: A Cadaveric Comparison of Intramedullary and Plate Fixation TechniquesJournal of Orthopaedic Research® · 2026
- Preclinical Cyclic Testing of Adjunctive Fixation of the Humeral Epicondyle in a Lateral Condylar Fracture Model: Ex Vivo Comparison of a Locking Plate versus a Light-curable Composite.Veterinary and comparative orthopaedics and traumatology : V.C.O.T · 2026
- Meniscus Allograft Interpositional Arthroplasty With Elastic Intramedullary Bioabsorbable Pin Fixation for Proximal Interphalangeal Joint Contracture.Hand (New York, N.Y.) · 2026
- Radial head arthroplasty restores axial construct stiffness in a Monteggia-like injury model: a biomechanical study.Archives of orthopaedic and trauma surgery · 2026
- Chondromyxoid Fibroma of the Metatarsals: Outcomes of En Bloc Resection and Intercalary Fibular Allograft Reconstruction in a Series of 9 Patients.Foot & ankle specialist · 2026
- Hybrid Fixation With a Low-Profile Locking Plate and Cannulated Compression Screw for Thumb Carpometacarpal Arthrodesis: A Retrospective Case Series.Annals of plastic surgery · 2026