Assessment of Scaphoid Screw Centrality Using a Dedicated Axial Fluoroscopic View: A CT-Validated Study.
Ilan Tzaytlin, Vadim Benkovich, Artsiom Abialevich, Layalee Abo-Naser, Guy Gershon, Nir Nazaro
Orthopedic Departemnt, Soroka Medical Center, Be'er Sheva, Israel. · Hand Surgery Unit, Orthopedic Department, Hadassah University Medical Center, Jerusalem, Israel.
Key points
From the paper's conclusion- The axial fluoroscopic view provides additional intraoperative information for assessing scaphoid screw position. Its value lies in detecting relative eccentricity at the guidewire or screw stage rather than replicating CT measurements. This technique may help surgeons judge case-specific acceptability of guidewire trajectory and reduce unrecognized malposition.
Abstract
BACKGROUND: Accurate intraoperative assessment of scaphoid screw position is limited by complex anatomy and standard fluoroscopic imaging. The surgical goal is near-central placement within a narrow corridor rather than a single geometric center. This study evaluates a dedicated axial fluoroscopic view for intraoperative assessment of screw centrality. METHODS: A retrospective study included 54 patients who underwent scaphoid fixation with intraoperative fluoroscopy, including a dedicated axial view, and postoperative computed tomography (CT). Screw deviation from the geometric center of the scaphoid waist was measured on CT and compared with axial view measurements. Agreement was assessed using Spearman correlation and Bland-Altman analysis. Subgroup analyses were performed based on fracture characteristics. RESULTS: Mean screw deviation was 1.73 ± 1.07 mm on CT and 1.25 ± 0.73 mm on the axial view. The mean absolute difference was 0.82 ± 0.99 mm. The axial view showed modest correlation with CT and a small systematic underestimation (mean bias -0.47 ± 1.20 mm), with wide 95% limits of agreement (-2.82 to 1.87 mm). Individual axial-view measurements should not be interpreted as interchangeable with CT. Displaced fractures demonstrated greater deviation on the axial view. Union was achieved in 93% of cases using a CT-based definition of osseous bridging. CONCLUSIONS: The axial fluoroscopic view provides additional intraoperative information for assessing scaphoid screw position. Its value lies in detecting relative eccentricity at the guidewire or screw stage rather than replicating CT measurements. This technique may help surgeons judge case-specific acceptability of guidewire trajectory and reduce unrecognized malposition.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An Algorithmic Approach to the Diagnosis and Staging of Scapholunate Instability in Distal Radius Fractures Using 7 Fluoroscopic Parameters Shows High Reliability and Accuracy.Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association · 2026
- Validity of fluoroscopic methods along the transsyndesmotic axis: A cadaveric study.Injury · 2026
- Avoiding pitfalls in percutaneous acetabular column screw placement using combined fluoroscopic check-point views: a technical note and case series.European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2026
- Influence of additional intra-articular arthroscopic visualization on surgeons' assessment of reduction adequacy in distal radius fractures: A survey-based observational study.Medicine · 2026
- Intraoperative 3D Tomosynthesis of the Elbow in a Simulated Prone-Position Setup: A Cadaveric Study.Diagnostics (Basel, Switzerland) · 2026
- A review of digital orthopedic techniques in pre- and intra-operative management of scaphoid fracture.EFORT open reviews · 2026