Analysis of Ablation Curves to Predict Transmural Lesions Using an Epicardial Bipolar Irrigated Radiofrequency Clamp in Pigs and Humans.
Luca Aerts, Julie Norup Hertel, Ben J M Hermans, Justin G L Luermans, Sevasti-Maria Chaldoupi, Ronald Lans, Arnela Saljic, Jos G Maessen,
Department of Cardiothoracic Surgery, Maastricht University Medical Centre, The Netherlands. · Department of Biomedical Sciences, University of Copenhagen, Denmark.
Key points
From the paper's conclusion- Three typical patterns associated with transmurality have been identified using the Gemini-S clamp. These preliminary types need to be tested and validated in the clinical setting during hybrid AF ablation.
Abstract
OBJECTIVE: It is currently not known at what time point a lesion made with the bipolar irrigated radiofrequency clamp (Cardioblate™ Gemini-S, Medtronic, Galway, Ireland) is transmural. This study aimed to explore typical patterns of ablation curves to potentially predict the transmurality of ablation lesions created with the Gemini-S clamp during atrial fibrillation (AF) ablation. METHODS: Ablations were performed on the atria and ventricles of 4 pigs. After ablation, the animals were sacrificed, and macroscopic evaluation was performed to identify the transmurality of the lesions. Typical patterns of ablation curves predicting transmurality were retrospectively identified. Second, a retrospective analysis of typical patterns in human ablation curves was performed to identify potential additional parameters associated with transmurality. Furthermore, the typical patterns identified in the pig experiments were retrospectively analyzed in human ablation curves. RESULTS: Three typical patterns of transmurality were identified in the ablation curves in pigs. Type I was a sharp impedance rise, type II was 3 power drops of at least a 2-energy-level decline in 1 ablation cycle, and type III was a power drop (<11 s) with subsequent hovering of power levels between 15 and 25 W. In 11 hybrid AF ablation patients, 272 applications were performed via the left (41.9%) and right (58.1%) side. Of these ablations, 46.3% were classified as transmural based on the 3 typical patterns of transmurality identified in the pig experiments. CONCLUSIONS: Three typical patterns associated with transmurality have been identified using the Gemini-S clamp. These preliminary types need to be tested and validated in the clinical setting during hybrid AF ablation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- An optimal output and electrode configuration for evaluating tissue transmurality - a preclinical study.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Clinical and experimental evaluation of averaged impedance drop during radiofrequency catheter ablation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2026
- Low-energy bipolar radiofrequency ablation using an inserted needle with a local temperature sensor: comparison of series and parallel electrical circuit connectionsEuropean Heart Journal · 2026
- A Novel Bipolar Radiofrequency Clamp for Efficient Isolation of the Left Atrial Posterior Wall.Innovations (Philadelphia, Pa.) · 2026
- Modified Unipolar Return Pulsed Field Ablation in Ventricular MyocardiumCirculation Arrhythmia and Electrophysiology · 2025
- Low Energy Bipolar Radiofrequency Ablation Combined With a Needle Electrode Connected in a Parallel Electrical Circuit to Create an Isolated Intramyocardial Lesion.Journal of cardiovascular electrophysiology · 2026