Clinical EEG and neuroscience

Automated Interictal Epileptiform Discharge Detector Scores in Serial EEG: Record-Level Concordance Did Not Extend to Within-Participant Change.

Murat Aydemir, Ezgi Tuna Erdoğan, Murat Mert Atmaca, Candan Gürses, Sacit Karamürsel

Department of Physiology, Koç University School of Medicine, Istanbul, Türkiye. · Department of Neurology, Sultan 2. Abdul Hamid Khan Educational and Research Hospital, Istanbul, Türkiye.

Observational · n=15

Key points

From the paper's conclusion
  • We tested whether three distinct pretrained detectors agreed on direction and rank ordering of within-participant change in serial EEG.MethodsThis secondary analysis included 45 routine EEG recordings from 15 adults with drug-resistant focal epilepsy at baseline, week 1, and month 3 in the active arm of a parent tDCS trial; treatment efficacy was not evaluated.

Abstract

ObjectiveAutomated deep learning-based interictal epileptiform discharge detectors have been validated for cross-sectional recognition, but longitudinal concordance is uncertain. We tested whether three distinct pretrained detectors agreed on direction and rank ordering of within-participant change in serial EEG.MethodsThis secondary analysis included 45 routine EEG recordings from 15 adults with drug-resistant focal epilepsy at baseline, week 1, and month 3 in the active arm of a parent tDCS trial; treatment efficacy was not evaluated. SpikeNet1, SpikeNet2, and Cyclops used pretrained weights and detector-specific pipelines on identically staged 25-min intervals. The record-level endpoint was the unthresholded mean window score. Concordance was characterized using record-level and change-score Spearman correlations, directional agreement, and a standardized detector × time repeated-measures ANOVA.ResultsDescriptive record-level correlations ranged from ρ = 0.341 to 0.696. Change-score correlations ranged from -0.011 to 0.379 (week 1) and from -0.089 to 0.536 (month 3). All six were smaller in absolute magnitude than their record-level counterparts and nonsignificant after Holm adjustment. Three-detector directional agreement was 7/15 (week 1) and 6/15 (month 3). The detector × time interaction was nonsignificant (Greenhouse-Geisser-corrected p = .142). Post hoc tolerance zones did not increase three-detector agreement. Thresholded SpikeNet2 and Cyclops outputs were sparse, and high-threshold agreement reflected shared zero change.ConclusionsRecord-level concordance did not extend to within-participant change or establish longitudinal interchangeability of mean window scores. Even for validated detectors, a longitudinal finding from one detector should not be assumed to generalize across detectors without independent corroboration.Parent trial registration: ClinicalTrials.gov (https://clinicaltrials.gov/study/NCT05951244), NCT05951244.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers