Continuous Glucose Monitoring Is Not Ready for the Diagnosis of Prediabetes.
J Hans DeVries
Profil, Neuss, Germany.
Key points
From the paper's conclusion- With the growing availability of over-the-counter continuous glucose monitors, the question arises whether Continuous Glucose Monitoring (CGM) can be used to diagnose prediabetes.
Abstract
With the growing availability of over-the-counter continuous glucose monitors, the question arises whether Continuous Glucose Monitoring (CGM) can be used to diagnose prediabetes. Traditionally, prediabetes is diagnosed using the oral Glucose Tolerance Test, a test with poor reproducibility. In a study from Singapore, there was large overlap between Ambulatory Glucose Profiles of people with prediabetes and normoglycemia. The optimal threshold of 5.64% Time above 140 mg/dL, the best discriminator, yielded 88% specificity and 48% sensitivity, indicating insufficient discriminatory value in many cases. In a study from Stanford, administration of a 75-gram oral glucose load while wearing CGM showed poor agreement with venous sampling between the category outcomes normoglycemia, prediabetes and diabetes. Future large cohort studies and AI may improve this situation. But for the moment, when a question of prediabetes arises, the most practical answer is to measure HbA1c, which allows for an elegant classification of normoglycemia, prediabetes and diabetes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Lack of Association Between Hemoglobin A1c and Continuous Glucose Monitor Metrics Among Individuals with Prediabetes and NormoglycemiaDiabetes Technology & Therapeutics · 2025
- Should continuous glucose monitoring redefine prediabetes?Diabetologia · 2026
- Continuous Glucose Monitoring in Prediabetes and Early Dysglycemia: A Narrative Review.Diabetes & metabolism journal · 2026
- Continuous glucose monitoring for diabetes prevention in adults with prediabetes: a scoping review.Frontiers in endocrinology · 2026
- The Case for Evaluating Continuous Glucose Monitoring as a Diagnostic Tool for Diabetes.Journal of diabetes science and technology · 2026
- Pharmacist-Led Continuous Glucose Monitoring for Prediabetes: Protocol for a Prospective Cohort Study.JMIR research protocols · 2026