The journal of vascular access

Transesophageal echocardiography-guided repositioning of a totally implantable venous access device that migrated into the left atrium across a patent foramen ovale.

William Amaya-Zúñiga, María Cifuentes-Ortiz, Daniel Benítez-Ávila, Roosevelt Fajardo-Gómez, Nohra Romero-Vanegas

Department of Anesthesia and Perioperative Medicine, Hospital Universitario Fundación Santa Fe de Bogotá, Bogotá, Colombia. · Department of Critical Care, Universidad del Rosario, Bogotá, Colombia.

Case report

Key points

From the paper's conclusion
  • We report the case of a patient whose catheter tip migrated across a patent foramen ovale into the left atrium.

Abstract

Migration of a totally implantable venous access port into the cardiac chambers can lead to a wide range of potentially serious complications. Malpositioned devices require thorough clinical assessment and timely intervention to minimize associated risks. We report the case of a patient whose catheter tip migrated across a patent foramen ovale into the left atrium. Because prompt surgical repositioning was required, conscious sedation and real-time transesophageal echocardiography were utilized to provide precise guidance. This approach enabled safe catheter withdrawal, accurate repositioning within the superior vena cava, and immediate confirmation of catheter function using a saline flush test.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers