What causes cup inclination outliers in primary total hip arthroplasty when using a digital inclinometer? A prospective cohort study.
Geert Meermans, Willem J Van Doorn, Jan-Jaap Kats, Moritz Innmann, George Grammatopoulos
Department of Orthopaedics, Bravis Hospital, Roosendaal, The Netherlands. · Department of Orthopaedics and Trauma Surgery, University of Heidelberg, Heidelberg, Germany.
Key points
From the paper's conclusion- Using a digital inclinometer allowed the surgeons to intraoperatively put the cup within 3° of the IOItarget in 93% of cases. This led to 74% of the acetabular components being placed within an RI target zone of 35°- 45°. The remaining outliers were caused by more of less than average deviation of the pelvic position at time of impaction.
Abstract
BACKGROUND: A suboptimal radiographic inclination (RI) angle of the cup has been associated with worse short- and long-term outcomes after primary total hip arthroplasty (THA). There are several factors that lead to RI outliers including cup version, pelvic orientation at time of cup impaction and angle of the cup introducer relative to the floor. The primary aim of this study was to analyse if a digital inclinometer helps achieving a target RI of cups in primary THA and what factors increase the risk of having a cup inclination outside the target zone. PATIENTS AND METHODS: In this prospective study, we included 200 consecutive patients undergoing primary THA. Preoperatively, the surgeon determined the target intraoperative inclination (IOI). The IOI was measured with the aid of a digital inclinometer after seating of the acetabular component. Anteroposterior pelvic radiographs were made to measure the RI of the acetabular component. The target zone for inclination was defined as 35-45° of RI. RESULTS: The mean RI was 37.9° SD 4.7. There were 53 outliers (26.5%) for the 35-45° zone. The mean absolute ΔIOIcup-IOItarget was 1.2° SD 1.0. The absolute ΔIOIcup-IOItarget was <1° in 108 patients (54%), <2° in 160 patients (80%) and <3° in 186 patients (93%). The absolute deviation from the mean ΔOImath-IOIcup, which corresponds with pelvic motion, was significantly higher in RI outliers compared with non-outliers for the 35-45° inclination zone (4.7°, SD 2.8 vs. 2.5°, SD 2.0 respectively) (p < 0.0001). CONCLUSIONS: Using a digital inclinometer allowed the surgeons to intraoperatively put the cup within 3° of the IOItarget in 93% of cases. This led to 74% of the acetabular components being placed within an RI target zone of 35°- 45°. The remaining outliers were caused by more of less than average deviation of the pelvic position at time of impaction.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Short-Term Clinicoradiological Outcomes of Primary Total Hip Arthroplasty With and Without Intraoperative Fluoroscopic Guidance: A Comparative Prospective Non-randomised Controlled (Quasi-experimental) Study.Journal of orthopaedic case reports · 2026
- Accuracy and risk factors for poor cup orientation in total hip arthroplasty: a comparison between supine and lateral registration using a portable navigation system.Archives of orthopaedic and trauma surgery · 2026
- Increased Accuracy of Acetabular Cup Positioning with an Automatic Impaction Device in Total Hip Arthroplasty.Clinics in orthopedic surgery · 2026
- Center of rotation restoration and acetabular cup sizing in total hip arthroplasty: the predictive role of radiographic femoral head diameter. A single-center study from the FP-UCBM hip study group.Frontiers in surgery · 2026
- Variations in Acetabular Offset Between the Preoperative Plan (mediCAD®) and Postoperative Radiographic Measurements: an overlooked source of error that affects restoration of global offset after THA.SICOT-J · 2026
- Radiographic outcomes of CT-based robotic-assisted versus manual direct anterior total hip arthroplasty with an acetabular-only registration workflow: a retrospective cohort study.Journal of robotic surgery · 2026