Prospective, Multicenter, Single-Arm, Open-Label Study of the Safety and Effectiveness of the Gel Stent in Angle Closure Glaucoma.
Michael Coote, Nathan Kerr, Paul Harasymowycz, Chelvin C A Sng, Katherine W X Lun, Yuen-Chieh Lee, Saumya Nagar, Daisy Yuan,
Royal Victorian Eye and Ear Hospital, Melbourne, Australia. · Bellevue Ophthalmology Clinics, Montreal, Canada.
Key points
From the paper's conclusion- Gel stent implantation had an acceptable safety profile and effectively lowered IOP and the medication burden in ACG.
Abstract
PURPOSE: To evaluate the gel stent's intraocular pressure (IOP)-lowering safety and effectiveness in angle closure glaucoma (ACG). METHODS: Prospective, multicenter, nonrandomized, 12-month, interventional study. Adults with ACG (iridotrabecular contact in ≥2 quadrants, glaucomatous optic disc damage, and visual field loss) and baseline IOP 20-35 mmHg received ab-interno gel stent implantation on Day 1. The primary endpoint was the proportion of patients achieving ≥20% IOP reduction from baseline without IOP-lowering medication increase at Month 12. RESULTS: Of 62 patients/eyes enrolled and implanted, 45 (72.6%) had a history of glaucoma procedures, 50 (80.6%) underwent ≥1 concomitant angle-widening procedure(s) (usually cataract surgery), and 59 (95.2%) completed the study. Baseline IOP was 24.0 mmHg on 2.3 IOP-lowering medications (mean). At Month 12, 80.5% (95% confidence interval [CI]: 68.4-88.8) of patients achieved the primary endpoint, with similar proportions in the presence (81.8%; 95% CI: 67.3-91.8) and absence (83.3%; 95% CI: 51.6-97.9) of concomitant angle-widening procedures. Month 12 mean changes from baseline in IOP and number of IOP-lowering medications were -8.9 mmHg (95% CI: -10.2 to -7.6) and -1.6 (95% CI: -1.9 to -1.2), respectively. Twenty-one eyes (33.9%) required needling. There were no surgical complications. Serious postoperative ocular adverse events occurred in 6 (9.7%) study eyes; all resolved without sequelae. CONCLUSION: Gel stent implantation had an acceptable safety profile and effectively lowered IOP and the medication burden in ACG.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Efficacy and safety of XEN63 gel implant with open conjunctiva technique.BMC ophthalmology · 2026
- The Adjunctive Effect of Goniotomy Combined with Phacoemulsification and Goniosynechialysis for Advanced Primary Angle-Closure Glaucoma with 360° Angle Closure: 12-Month Outcomes.Clinical ophthalmology (Auckland, N.Z.) · 2026
- Safety and Effectiveness Outcomes of Ab Interno Canaloplasty in Open-angle Glaucoma: 10-year ResultsJOURNAL OF CURRENT GLAUCOMA PRACTICE · 2026
- Safety and Effectiveness of a Gel Stent in Refractory Glaucoma in China: Real-World Evidence at 12 Months.Clinical ophthalmology (Auckland, N.Z.) · 2026
- One-year outcomes of combined phacoemulsification and goniosynechialysis in primary angle-closure glaucoma with cataracts: a multicenter prospective study.Eye and vision (London, England) · 2026
- Three-year outcomes of combined surgical peripheral iridectomy, goniosynechialysis and goniotomy for advanced primary angle-closure glaucoma: a multicentre study.The British journal of ophthalmology · 2026