Establishment of Reference Ranges for Coagulation Factors and Anticoagulant Protein Activities in Healthy Chinese Adults.
Juhua Dai, Qian Song, Liuqing Yang, Xiaohui Liu, Xinping Sun
Department of Laboratory Medicine, Peking University International Hospital, Beijing, China. · Department of Laboratory Medicine, Xingtai General Hospital, North China Medical and Health Group, Xingtai, China.
Key points
From the paper's conclusion- A sex difference remained for PS after BH adjustment (q = 2.13 × 10-9); the male and female RIs were 73.1-139.8% and 58.7-117.6%, respectively.ConclusionWe established platform-specific RIs for eight coagulation-factor and three anticoagulant-protein activities in healthy Chinese adults sampled at three centers using the Mindray CX-9000 coagulation system.
Abstract
ObjectiveThis study aimed to establish reference intervals (RIs) for the functional activities of major coagulation factors and anticoagulant proteins in healthy Chinese adults.MethodsThis multicenter cross-sectional study included eligible adults from three centers. Standard sodium citrate-anticoagulated plasma samples were tested via a Mindray fully automated coagulation analyzer. After Dixon screening, which excluded no observations, central 95% RIs were estimated uniformly using the nonparametric 2.5th-97.5th percentiles, with 90% confidence intervals (CIs) for both limits. Sex differences and age associations were assessed using the Mann-Whitney U test and Spearman correlation, respectively, with Benjamini-Hochberg (BH) adjustment. Center effects were evaluated after adjustment for age and sex.ResultsValid sample sizes ranged 289-448, with RIs generated for all analytes. The overall reference ranges of coagulation factor activity (%) were as follows: FII, 68.0-119.2; FV, 63.9-122.5; FVII, 65.2-125.8; FVIII, 58.1-157.5; FIX, 70.5-152.1; FX, 63.7-117.6; FXI, 63.4-149.6; FXII, 45.3-145.1. The reference ranges of anticoagulant protein activity (%) were protein S (PS), 60.2-136.8; protein C (PC), 68.1-143.1; and antithrombin (AT), 80.4-120.9. A sex difference remained for PS after BH adjustment (q = 2.13 × 10-9); the male and female RIs were 73.1-139.8% and 58.7-117.6%, respectively.ConclusionWe established platform-specific RIs for eight coagulation-factor and three anticoagulant-protein activities in healthy Chinese adults sampled at three centers using the Mindray CX-9000 coagulation system. PS sex-specific estimates are provided for local verification. These reference limits are specific to the sampled populations and analytical system while local laboratory verification is required before clinical implementation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Instrument-specific reference intervals for coagulation factor XII and antithrombin III: A regional study in Jiangsu, China.Journal of immunological methods · 2026
- Age-related Global Coagulation Changes: From Neonate to Old Age.Seminars in thrombosis and hemostasis · 2026
- 8 Indirect Estimation of Age-Specific Pediatric Activated Partial Thromboplastin Time Reference IntervalsAmerican Journal of Clinical Pathology · 2026
- Impact of reference interval source on result classification in pediatric coagulation testing.Clinica chimica acta; international journal of clinical chemistry · 2026
- A-258 Heterogeneity of the Anticoagulant Profile in Carriers of Thrombophilic Mutations: A Laboratory-Based Cluster AnalysisClinical Chemistry · 2026
- Diagnostic challenges in current laboratory practice at Italian Hemophilia Centers: findings from a nationwide surveyBleeding Thrombosis and Vascular Biology · 2026