Accuracy of Computer-Assisted Navigation in Total Ankle Arthroplasty Surgery Using Computed Tomography: A Short Report.
Taylor R Rakauskas, Adam P Henderson, Gabrielle Meli, Cole Herbel, Seif El Masry, François Boux De Casson, Cyril Hamad, Edward Haupt
Mayo Clinic Florida, Jacksonville, FL, USA. · Blue Ortho, Meylan, Grenoble, France.
Key points
From the paper's conclusion- Although discrepancies persisted in tibial rotational alignment and talar positioning, the low variability across measurements suggests that CAN may support accurate implant positioning. These short-term positioning accuracy findings support the utility of CAN as a potential tool for enhancing implant positioning reproducibility in TAA.
Abstract
BACKGROUND: Computer-assisted navigation (CAN) offers dynamic intraoperative guidance in total ankle arthroplasty (TAA). Although laboratory data suggest superior precision, in vivo evidence is lacking. This study used weightbearing computed tomography (WBCT) to evaluate CAN accuracy of postoperative TAA implant position relative to the plan. We hypothesized that absolute discrepancies would be <3 mm or <3°. METHODS: Thirteen consecutive patients (November 2024-November 2025) underwent TAA using CAN. Preoperative computed tomography-based 3D models were used for templating. Postoperative WBCT scans were used to measure final implant seating, and computerized measurements were used to identify implant placement discrepancy. We analyzed absolute discrepancies between planned vs final postoperative positions. RESULTS: Tibial absolute discrepancies were within desired limits for varus/valgus (1.65° ± 1.43°), slope (1.87° ± 1.34°), and cut height (0.68 ± 0.62 mm) but not for axial rotation (3.97 ± 2.64). For the talus, varus/valgus (2.93° ± 1.86°) and cut height (0.66 ± 0.71 mm) absolute discrepancies were within desired limits. The increased talar slope deviation (3.90 ± 2.89) in the final position may be due to non-navigated implant impaction. CONCLUSION: Although discrepancies persisted in tibial rotational alignment and talar positioning, the low variability across measurements suggests that CAN may support accurate implant positioning. These short-term positioning accuracy findings support the utility of CAN as a potential tool for enhancing implant positioning reproducibility in TAA. LEVEL OF EVIDENCE: Level IV, case series.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Accuracy and Precision of Computer-assisted Navigation Compared With Patient-Specific Instrumentation and Conventional Methods in Total Ankle Arthroplasty.The Journal of the American Academy of Orthopaedic Surgeons · 2026
- Accuracy of implant positioning in total ankle arthroplasty using patient-specific instrumentation: A weightbearing CT analysis of landmark-based parameters.Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons · 2026
- Variability of Visual Assessment in Determining Tibial Component Alignment in Total Ankle Arthroplasty.Foot & ankle orthopaedics · 2026
- Association Between Dynamic Tibiotalar Motion and Patient-Reported Outcomes After Total Ankle Arthroplasty With a Total Talar Prosthesis.Foot & ankle orthopaedics · 2026
- Computer-Navigated Versus Conventional Total Knee Arthroplasty Using the Same VEGA Prosthesis: A Retrospective Cohort StudyProsthesis · 2026
- Tibial Component Malalignment in Total Ankle Arthroplasty.Medicina (Kaunas, Lithuania) · 2026