Obesity Is Associated With Earlier Mechanical Failure Following Total Ankle Arthroplasty Independent of Tibial Stem or Keel Augmentation.
Corinne Sommi, Grant M Thomas, Joseph A S McCahon, Thomas Dowd, Anny Hsu, Selene G Parekh, David I Pedowitz
Rothman Orthopaedic Institute at Thomas Jefferson University, Philadelphia, PA, USA. · NYU Langone Orthopaedics, New York, NY, USA.
Key points
From the paper's conclusion- After adjusting for age, implant design, and subtalar fusion, BMI ≥30 was independently associated with a significantly higher hazard of mechanical failure and shorter time to revision. Despite the proposed biomechanical advantages of increased tibial fixation, the stemmed or keeled implants used in this cohort were not associated with improved survivorship or patient-reported outcomes at mid-term follow-up.
Abstract
BACKGROUND: Obesity and elevated body mass have been associated with increased rates of aseptic loosening and implant failure following total ankle arthroplasty (TAA). The purpose of this study was to investigate the role for primary stemmed or keeled tibial implants as a protective factor against aseptic loosening in this patient population. METHODS: We performed a retrospective cohort study of 699 modern primary TAAs with a minimum 2-year follow-up performed at a large academic institution between 2010 and 2023. Patients were stratified by body mass index (BMI <30 vs ≥30) and implant design (stem/keel vs low-profile). The primary outcome was mechanical failure requiring revision of metal components. Secondary outcomes included patient-reported outcomes (FAAM-ADL and VAS). Kaplan-Meier analysis and multivariable cox proportional hazards were used to assess revision risk. RESULTS: Of 699 primary TAAs, 39 (5.6%) experienced mechanical failure at a mean follow-up of 5.6 years. After adjustment for age, implant type, and subtalar fusion, BMI ≥30 was associated with a greater than 2-fold increased hazard of mechanical failure (HR 2.27, 95% CI 1.07-4.81, P = .033). Kaplan-Meier survivorship was similar for low-profile (93.7%) and stem/keel (96.1%) implants at 5 years. When stratified by both BMI and implant type, with the numbers available, no significant difference could be detected in mechanical failure, time to revision, or patient-reported outcomes were observed (P > .05). Obese patients had worse preoperative FAAM-ADL and VAS scores but achieved comparable postoperative outcomes. CONCLUSION: After adjusting for age, implant design, and subtalar fusion, BMI ≥30 was independently associated with a significantly higher hazard of mechanical failure and shorter time to revision. Despite the proposed biomechanical advantages of increased tibial fixation, the stemmed or keeled implants used in this cohort were not associated with improved survivorship or patient-reported outcomes at mid-term follow-up. LEVEL OF EVIDENCE: Level III, retrospective cohort study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Restricting Access to Care by Using Body Mass Index as Eligibility Criteria for Total Ankle Arthroplasty: A Retrospective Cohort StudyFoot & Ankle International · 2026
- Ankle Arthrodesis Associated With Risk of Progression to Subtalar Arthrodesis Compared to Total Ankle Arthroplasty.Foot & ankle orthopaedics · 2026
- Association between body mass index and early clinical efficacy and complications after total knee arthroplasty.Journal of orthopaedic surgery (Hong Kong) · 2026
- Is There a Need for a Central Stem within the Keel of Primary Tibial Baseplates in Total Knee Arthroplasty.The journal of knee surgery · 2026
- Implant Choice Impacts Initial Tibial Component Micromotion in Total Ankle Replacement: A Biomechanical Analysis.The Journal of bone and joint surgery. American volume · 2026
- Tibial implant migration of cemented and cementless total and unicompartmental knee arthroplasties in patients with low and normal bone mineral density: a 5-year RSA cohort study of 397 cases.Acta orthopaedica · 2026